Clinical Applications for Spread of Excitation Functions Obtained Via Electrically Evoked Compound Action Potentials
Katelyn A Berg1, Andrea J DeFreese1, Allyson L Sisler-Dinwiddie1
1Department of Hearing and Speech Sciences, Vanderbilt University Medical Center, Nashville, Tennessee.
Objectives:
To assess the clinical utility of spread of excitation (SOE) functions obtained via electrically evoked compound action potentials (eCAP) to 1) identify electrode array tip fold-over, 2) predict electrode placement factors confirmed via postoperative computed tomography (CT) imaging, and 3) predict postoperative speech recognition through the first year post-activation in a large clinical sample.
Study Design:
Retrospective case review.
Setting:
Cochlear implant (CI) program at a tertiary medical center.
Patients:
Two hundred seventy-two ears (238 patients) with Cochlear Ltd. CIs (mean age = 46 yr, range = 9 mo-93 yr, 50% female) implanted between August 2014 and December 2022 were included.
Main Outcome Measures:
eCAP SOE widths (mm) (probe electrodes 5, 11, and 17), incidence of electrode tip fold-over, CT imaging data (electrode-to-modiolus distance, angular insertion depth, scalar location), and speech recognition outcomes (consonant-nucleus-consonant [CNC], AzBio quiet, and +5 dB SNR) through the first year after CI activation.
Results:
1) eCAP SOE demonstrated a sensitivity of 85.7% for identifying tip fold-over instances that were confirmed by CT imaging. In the current dataset, the tip fold-over incidence rate was 3.1% (7 patients), with all instances involving a precurved electrode array. 2) There was a significant positive relationship between eCAP SOE and mean electrode-to-modiolus distance for precurved arrays, and a significant positive relationship between eCAP SOE and angular insertion depth for straight arrays. No relationships between eCAP SOE and scalar location or cochlea diameter were found in this sample. 3) There were no significant relationships between eCAP SOE and speech recognition outcomes for any measure or time point, except for a weak negative correlation between average eCAP SOE widths and CNC word scores at 6 months post-activation for precurved arrays.
Conclusions:
In the absence of intraoperative CT or fluoroscopic imaging, eCAP SOE is a reasonable alternative method for identifying electrode array tip fold-over and should be routinely measured intraoperatively, especially for precurved electrode arrays with a sheath.
More Related Videos
08:32External Excitation of Neurons Using Electric and Magnetic Fields in One- and Two-dimensional Cultures
Published on: May 7, 2017
06:55Author Spotlight: Advancing Spinal Cord Stimulation - Exploring the Cellular Responses of Motor Neurons Through Patch-Clamp Electrophysiology
Published on: September 8, 2023
Related Concept Videos
Action Potential
Membrane potential in neurons
Neurons typically have a resting membrane potential of about -70 millivolts (mV). When they...
Action Potential: Phases of Stimulation
Resting Phase:
In this phase, the cell's membrane is at its resting potential, typically around -70 millivolts (mV) for neurons. Inside the cell, there is a higher concentration of potassium ions (K+) and a lower concentration of sodium ions (Na+). Voltage-gated sodium channels are closed, and...
Postsynaptic Potential (PSP)
There are two types of receptors: ionotropic and metabotropic.
The ionotropic receptor is the membrane protein that has an...
Graded Potential
Graded potentials fall into two categories: depolarizing and hyperpolarizing. Depolarizing graded potentials typically occur when sodium (Na+) or...
Action Potentials