Echo reference standards in ophthalmic ultrasonography
Ultrasound in Medicine & Biology
|January 1, 1985
Summary
This study introduces the W38 Hema reflector, a novel standard for ophthalmic ultrasonography. Its material properties ensure accurate calibration of ultrasonic instruments, improving diagnostic reliability.
Area of Science:
- Ophthalmic ultrasonography
- Biomedical acoustics
- Medical device calibration
Background:
- Standardization in ophthalmic ultrasonography is crucial for accurate diagnostic measurements.
- Current echo standards may lack the required properties for precise calibration.
- The W38 Hema reflector is proposed as a potential solution.
Purpose of the Study:
- To determine and verify the acoustical properties of the W38 Hema reflector.
- To assess its suitability as a plane echo interface for ophthalmic ultrasonography.
- To demonstrate its application in instrument performance control and calibration.
Main Methods:
- Acoustical property determination of the W38 Hema reflector using soft corneal contact lens material.
- Independent verification measurements.
- Development of a conversion diagram for existing echo references.
- Application demonstration in performance control and calibration.
Main Results:
- The W38 Hema reflector material meets the requirements for plane echo interfaces.
- The material is suitable for use as an echo standard in ophthalmic ultrasonography.
- A conversion diagram facilitates the transition between different echo references.
- Successful application in controlling and calibrating ultrasonic instruments.
Conclusions:
- The W38 Hema reflector is a viable and effective echo standard for ophthalmic ultrasonography.
- Its use enhances the accuracy and reliability of ultrasonic measurements.
- This development aids in the precise calibration and performance control of ophthalmic ultrasound devices.