A Novel Noninvasive Screening Tool for Dry Eye Disease
Sabrina Vaccaro1, Massimiliano Borselli1, Giovanni Scalia1
1Department of Ophthalmology, University Magna Græcia of Catanzaro, Viale Europa, 88100 Catanzaro, Italy.
Diagnostics (Basel, Switzerland)
|June 27, 2024
Summary
The new DEvice© tool accurately screens for dry eye disease (DED) by measuring ocular surface relative humidity (RH). This noninvasive method shows promise for rapid DED diagnosis in clinical settings.
Area of Science:
- Ophthalmology
- Medical Devices
- Diagnostic Tools
Background:
- Dry eye disease (DED) is a prevalent condition affecting ocular surface health.
- Accurate and efficient screening tools are crucial for timely DED diagnosis and management.
Purpose of the Study:
- To evaluate the feasibility and diagnostic accuracy of the novel DEvice© tool for DED screening.
- To assess the utility of measuring ocular surface relative humidity (RH) and temperature for DED detection.
Main Methods:
- A comparative study involving 40 patients, classified as DED or non-DED using the Keratograph 5M.
- DEvice© was used to measure ocular surface RH and temperature, alongside OSDI symptom scoring.
- Key metrics included noninvasive keratograph breakup time (NIKBUT), tear meniscus height (TMH), meibomian gland loss (MGL), and bulbar redness.
Main Results:
- DEvice© detected significantly higher ocular surface RH in DED patients compared to controls (85.93% vs. 73.05%, p=0.0049).
- A significant correlation was observed between RH and OSDI scores (r=0.406, p=0.009).
- The RH parameter demonstrated a discriminating power for DED with an AUC of 0.782 (p=0.0022).
Conclusions:
- The DEvice© tool effectively differentiates DED patients from healthy individuals.
- Ocular surface RH measurement using DEvice© shows high sensitivity, making it suitable for fast, noninvasive DED screening.