Clinical Validation of Carotid-Femoral Pulse Wave Velocity Measurement Using a Multi-Beam Laser Vibrometer: The
Smriti Badhwar1, Louise Marais1, Hakim Khettab2
1Paris Cardiovascular Research Center (PARCC) Institut National de la Santé Et de la Researche Médicale (INSERM), Paris, France (S.B., L.M., F.P., R.M.B., P.B.).
Background:
Carotid-femoral pulse wave velocity (cfPWV) is the gold standard for noninvasive arterial stiffness assessment, an independent predictor of cardiovascular disease, and a potential parameter to guide therapy. However, cfPWV is not routinely measured in clinical practice due to the unavailability of a low-cost, operator-friendly, and independent device. The current study validated a novel laser Doppler vibrometry (LDV)-based measurement of cfPWV against the reference technique.
Methods:
In 100 (50 men) hypertensive patients, cfPWV was measured using applanation tonometry (Sphygmocor) and the novel LDV device. This device has 2 handpieces with 6 laser beams each that simultaneously measure vibrations from the skin surface at carotid and femoral sites. Pulse wave velocity is calculated using ECG for the identification of cardiac cycles. An ECG-independent method was also devised. Cardiovascular risk score was calculated for patients between 40 and 75 years old using the WHO risk scoring chart.
Results:
LDV-based cfPWV correlated significantly with tonometry (r=0.86, P<0.0001 ECG-dependent [cfPWVLDV_ECG] and r=0.80, P<0.001 ECG-independent [cfPWVLDV_w/oECG] methods). Bland-Altman analysis showed nonsignificant bias (0.65 m/s) and acceptable SD (1.27 m/s) between methods. Intraobserver coefficient of variance for LDV was 4.7% (95% CI, 3.0%-5.5%), and interobserver coefficient of variance was 5.87%. CfPWV correlated significantly with CVD risk (r=0.64, P<0.001; r=0.41, P=0.003; and r=0.37, P=0.006 for tonometry, LDV-with, and LDV-without ECG, respectively).
Conclusions:
The study demonstrates clinical validity of the LDV device. The LDV provides a simple, noninvasive, operator-independent method to measure cfPWV for assessing arterial stiffness, comparable to the standard existing techniques.
Registration:
URL: https://clinicaltrials.gov/study/NCT03446430; Unique identifier: NCT03446430.
Related Concept Videos
Assessment of blood pressure in brachial artery(two-step method)
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Pulse
The pulse serves as a clinical...
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment: