Related Experiment Video
Updated: Jun 22, 2025

Non-fluoroscopic Catheter Tracking for Fluoroscopy Reduction in Interventional Electrophysiology
Published on: May 26, 2015
Impact of Defibrillator Electrode Placement on Outcome of Electrical Cardioversion of Atrial Fibrillation: A Pilot
Luai Madanat1, Josh France2, Nolan Shoukri2
1Department of Cardiovascular Medicine, William Beaumont University Hospital Corewell Health East Royal Oak MI USA.
Background:
Anterior-posterior electrode placement is preferred in electrical cardioversion of atrial fibrillation. However, the optimal anterior-posterior electrode position in relation to the heart is not studied.
Methods And Results:
We performed a prospective observational study on patients presenting for cardioversion of atrial fibrillation. Electrodes were placed in the anterior-posterior position and shock was delivered in a step-up approach (100 J→200 J→360 J). Fluoroscopic images were obtained, and distances were measured from points A, midanterior electrode; and B, midposterior electrode, to midpoint of the cardiac silhouette. Patients requiring one 100 J shock for cardioversion success (group I) were compared with those requiring >1 shock/100 J (group II). Logistic regression was used to determine the impact of electrode distance on low energy (100 J) cardioversion success. Computed tomography scans from this cohort were analyzed for anatomic landmark correlation to the cardiac silhouette. Of the 87 patients included, 54 (62%) comprised group I and 33 (38%) group II. Group I had significantly lower distances from the mid-cardiac silhouette to points A (5.0±2.4 versus 7.4±3.3 cm; P<0.001) and B (7.3±3.0 versus 10.0±3.8 cm; P=0.002) compared with group II. On multivariate analysis, higher distances from the mid-cardiac silhouette to point A (odds ratio, 1.33 [95% CI, 1.07-1.70]; P=0.01) and B (odds rsatio, 1.24 [95% CI, 1.05-1.50]; P=0.01) were independent predictors of low energy (100 J) cardioversion failure. Based on review of computed tomography scans, we suggest that the xiphoid process may be an easy landmark to guide proximity to the myocardium.
Conclusions:
In anterior-posterior electrode placement, closer proximity to the cardiac silhouette predicts successful 100 J cardioversion irrespective of clinical factors.
More Related Videos
05:43Author Spotlight: Improving Lesion Contiguity in Pulmonary Vein Isolation via Proactive Esophageal Cooling
Published on: April 19, 2024
16:40A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
Related Concept Videos
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin...
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow...