The New Semisynthetic TORP: A Prosthesis for Ossicular Reconstruction Both With the Absence and the Presence of the
Giuseppe Malafronte1, Anna Rita Fetoni2, Gaetano Motta3
1Department of Otolaryngology Head and Neck Surgery, San Giuseppe Moscati Hospital, Avellino, Italy.
Objectives:
To describe the new semisynthetic total ossicular replacement prosthesis (New-SSTORP) and to evaluate the New-SSTORP ossiculoplasty results both with the presence and absence of the stapes superstructure.
Study Design:
Prospective study.
Setting:
Tertiary referral center.
Methods:
From April 2023 to May 2023, 18 New-SSTORP ossiculoplasties were performed by the first author. In all patients, the New-SSTORP was interposed between the footplate and the eardrum. The study group was divided into two groups (group A and group B). Group A included 13 patients with the absence of stapes superstructure. Group B included five patients with the presence of stapes superstructure. A successful reconstruction was defined as a postoperative air-bone gap ABG ≤20 dB. For all patients of groups A and B, the last audiometric control considered was performed in January 2024. The mean follow-up was 8½ months. The χ2 test was used to compare results. p < 0.05 was considered significant.
Main Outcome Measures:
Mean postoperative ABG ≤20 dB.
Results:
At the end of follow-up, the overall success rate (ABG ≤20 dB) of New-SSTORP ossiculoplasty was obtained in 88.8% (n = 16 of 18) of cases. In group A, the success rate of New-SSTORP ossiculoplasty occurred in 84.6% (n = 11 of 13) of cases, and in group B, the success rate of New-SSTORP ossiculoplasty occurred in 100% (n = 5 of 5) of cases. There was no audiological statistically significant difference between groups A and B (Fisher value is 1; p < 0.05). In all cases, the time for positioning of New-SSTORP was about 5 minutes.
Conclusion:
The New-SSTORP has a minimal technical challenge for building and placement. The New-SSTORP ossiculoplasty results are very good both with the presence and absence of SS.