Acquisition Efficiency and Technical Repeatability of Dual-Frequency 3D Vector MR Elastography of the Liver
Caixin Qiu1,2, Kevin J Glaser2, Nana Owusu2
1Department of Radiology, Tianjin First Central Hospital, Tianjin, China.
Background:
MR elastography (MRE) at 60 Hz is widely used for staging liver fibrosis. MRE with lower frequencies may provide inflammation biomarkers.
Purpose:
To establish a practical simultaneous dual-frequency liver MRE protocol at both 30 Hz and 60 Hz during a single examination and validate the occurrence of second harmonic waves at 30 Hz.
Study Type:
Retrospective.
Subjects:
One hundred six patients (48 females, age: 50.0 ± 13.4 years) were divided as follows: Cohort One (15 patients with chronic liver disease [CLD] and 25 healthy volunteers) with simultaneous dual-frequency MRE. Cohort Two (66 patients with CLD) with second harmonic MRE.
Field Strength/Sequence:
3-T, single- or dual-frequency MRE at 30 Hz and 60 Hz.
Assessment:
Liver stiffness (LS) in both cohorts was evaluated with manually placed volumetric ROIs by two independent analyzers. Image quality was assessed by three independent readers on a 4-point scale (0-3: none/failed, fair, moderate, excellent) based on the depth of wave propagation with 1/3 incremental penetration. The success rate was derived from the percentage of nonzero quality scores.
Statistical Tests:
Measurement agreement, bias, and repeatability of LS were assessed using intraclass correlation coefficients (ICCs), Bland-Altman plots, and repeatability coefficient (RC). Mann-Whitney U tests were used to evaluate the differences in image quality between different methods. A P-value <0.05 was considered statistically significant.
Results:
Success rate was 97.5% in Cohort One and 91% success rate for the second harmonic MRE in Cohort Two. The second harmonic and conventional MRE showed excellent agreement in LS (all ICCs >0.90). The quality scores for the second harmonic wave images were lower than those from the conventional MRE (Z = -4.523).
Data Conclusion:
Compared with conventional and second harmonic methods, simultaneous dual-frequency had better image quality, high success rate and the advantage of intrinsic co-registration, while the second harmonic method can be an alternative if custom waveform is not available.
Evidence Level:
3 TECHNICAL EFFICACY: Stage 1.