Novel Pulse Waveform Index by Ambulatory Blood Pressure Monitoring and Cardiac Function: A Pilot Study
Keisuke Narita1, Zihan Yuan2, Nobuhiko Yasui2
1Division of Cardiovascular Medicine, Department of Internal Medicine, Jichi Medical University School of Medicine, Shimotsuke, Japan.
Background:
A simple ambulatory measure of cardiac function could be helpful for monitoring heart failure patients.
Objectives:
The purpose of this paper was to determine whether a novel pulse waveform analysis using data obtained by our developed multisensor-ambulatory blood pressure monitoring (ABPM) device, the 'Sf/Am' ratio, is associated with echocardiographic left ventricular ejection fraction (LVEF).
Methods:
Multisensor-ABPM was conducted twice at baseline in 20 heart failure (HF) patients with HF-reduced LVEF or HF-preserved LVEF (median age 66 years, male 65%) and over a 6- to 12-month follow-up after patient-tailored treatment. We assessed the changes in the pulse waveform index Sf/Am and LVEF that occurred between the baseline and follow-up. The Sf/Am consists of the area of the ejection part in the square forward wave (Sf) and the amplitude of the measured wave (Am). We divided the patients into the recovered (n = 11) and not-recovered (n = 9) groups defined by a ≥10% increase in LVEF.
Results:
Although the ambulatory BP levels and variabilities did not change in either group, the Sf/Am increased significantly in the recovered group (baseline 21.4 ± 4.5; follow-up, 25.6 ± 3.7, P = 0.004). The not-recovered group showed no difference between the baseline and follow-up. The follow-up/baseline Sf/Am ratio was significantly associated with the LVEF ratio (r = 0.469, P = 0.037). The Sf/Am was significantly correlated with the LVEF in overall measurements (n = 40, r = 0.491, P = 0.001).
Conclusions:
These results demonstrated that a novel noninvasive pulse waveform index, the Sf/Am measured by multisensor-ABPM is associated with LVEF. The Sf/Am may be useful for estimating cardiac function.
Related Concept Videos
Pulse amplitude and quality
A weak or absent pulse may indicate reduced cardiac output or poor left ventricular contraction, which can be signs of cardiovascular dysfunction or...
Pulse
The pulse serves as a clinical...
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Pulse Assessment Sites
Special considerations while measuring pulse
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...