Mimicking and Measuring Occlusal Erosive Tooth Wear with the "Rub&Roll" and Non-contact Profilometry
Published on: February 2, 2018
Contact point marking with the OccluSense system-an in vitro study on reliability and validity
Stefanie Popa1,2, M Oliver Ahlers1,2
1Department of Prosthetic Dentistry, Center for Dental and Oral Medicine, University Medical Center Hamburg-Eppendorf, Hamburg, Germany.
OccluSense offers reliable and valid occlusal contact marking, comparable to traditional methods when using a modified evaluation strategy. This hybrid system
Area of Science:
- Dental materials science
- Occlusion analysis
- Diagnostic technology
Background:
- Occlusal indicators are used to visualize occlusal contacts.
- OccluSense is a hybrid system combining analogue contact marking and digital registration.
- The reliability and validity of OccluSense's contact marking require validation.
Purpose of the Study:
- To evaluate the reliability and validity of OccluSense analogue contact staining.
- To compare OccluSense with Arti-Fol and the reference standard Gnatho-Film.
- To assess the accuracy of occlusal contact registration.
Main Methods:
- An in-vitro study simulated static occlusion using a custom apparatus.
- Occlusal contacts were stained 50 times with Gnatho-Film to establish a baseline.
- Contacts were then stained with Arti-Fol or OccluSense (including a modified evaluation) and compared to the Gnatho-Film baseline.
Main Results:
- OccluSense registered the same number of contacts in over 94% of cases across both evaluation methods.
- Neither OccluSense nor Arti-Fol consistently matched the exact number of contacts registered by Gnatho-Film.
- The modified evaluation strategy for OccluSense (OccluSense (mod.)) significantly improved validity results.
Conclusions:
- OccluSense, with a modified evaluation strategy, demonstrates reliability and validity comparable to the reference standard.
- Pre-coloring with Gnatho-Film and the modified evaluation may become a future standard for OccluSense analysis.
- This approach enhances the analysis of occlusal contacts.