Real-time lung weight measurement during clinical ex vivo lung perfusion
Ichiro Sakanoue1, Toshihiro Okamoto2, Kamal S Ayyat1
1Department of Thoracic and Cardiovascular Surgery, Cleveland Clinic, Cleveland, Ohio; Department of Inflammation and Immunity, Lerner Research Institute, Cleveland Clinic, Cleveland, Ohio.
Background:
Real-time lung weight (LW) measurement is a simple and noninvasive technique for detecting extravascular lung water during ex vivo lung perfusion (EVLP). We investigated the feasibility of real-time LW measurement in clinical EVLP as a predictor of transplant suitability and post-transplant outcomes.
Methods:
In our clinical acellular EVLP protocol, real-time LW was measured in 117 EVLP cases from June 2019 to June 2022. The estimated LW gain at each time point was calculated using a scale placed under the organ chamber. The lungs were classified into 4 categories based on LW adjusted for height and compared between suitable and unsuitable cases. The relationship between estimated LW gain and primary graft dysfunction was also investigated.
Results:
The estimated LW gain during the EVLP significantly correlated with the LW gain (post-EVLP LW and pre-EVLP LW) measured on the back table (R2 = 0.61, p < 0.01). In the adjusted LW categories 2 to 4, the estimated LW gain at 0-1 hour after EVLP was significantly higher in unsuitable cases than in suitable cases. The area under the curve for the estimated LW gain was ≥0.80. Primary graft dysfunction grades 0 to 1 had a significantly lower estimated LW gain at 60 minutes than grades 2 to 3 (-43 vs 1 g, p < 0.01).
Conclusions:
Real-time lung measurements can predict transplant suitability and post-transplant outcomes by the early detection of extravascular lung water during the initial 1 hour of EVLP.