Evaluation of the performance of digital x-ray systems in pelvis radiography
Antonio Jreije1,2, Leonid Krynke1, Birutė Gricienė1,3
1Vilnius University Hospital Santaros Klinikos, Vilnius, Lithuania.
Abstract:
The aim of this study was to investigate the performance of eight digital radiography systems and to optimise the dose-image quality relationship for digital pelvis radiography. The study involved eight digital radiography systems used for general examinations at Vilnius University Hospital Santaros Klinikos. An anthropomorphic pelvic phantom (CIRS, US) was used to simulate a patient undergoing clinical pelvis radiography. Dose quantities entrance surface dose, dose area product (DAP) and exposure parameters (kVp, mA, mAs) were measured and the effects on the images were evaluated, considering physical contrast to noise ratio (CNR) and observer-based evaluations as image quality metrics. Increasing the tube voltage by 5 kVp from standard protocol led to a reduction in radiation dose (DAP) by 12%-20% with a slight impact on image quality (CNR decreases by 2%-10%). There was an inter-observer variability in image rating across different equipment (kappa value between 0 and 0.3); however, both observers agreed that increasing kVp up to 85-90 kV had no effect on perceived image quality. The results indicate that optimisation strategies should be tailored specifically for each x-ray system since significant performance differences and wide variations in radiation dose exist across various digital radiography systems used in clinical settings. The use of high kVp can be used for dose optimisation in digital pelvis radiography without compromising image diagnostic accuracy.
Related Concept Videos
X-ray Imaging
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Radiological Investigation I: X-ray and CT
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Ultrasonography
During an ultrasonography procedure, a handheld device called...