Instrumented Mouthguard Decoupling Affects Measured Head Kinematic Accuracy.
Ryan A Gellner1, Mark T Begonia2, Matthew Wood2
1Virginia Tech (Biomedical Engineering and Mechanics), Blacksburg, VA, USA. gryan3@vt.edu.
Annals of Biomedical Engineering
|July 2, 2024
Summary
Instrumented mouthguards can inaccurately measure head impacts due to decoupling. Larger gaps and frontal impacts significantly increase measurement error, requiring researchers to consider these limitations.
Area of Science:
- Biomechanics
- Sports Science
- Injury Prevention
Background:
- Boil-and-bite instrumented mouthguards are used to measure head kinematics in sports impacts.
- These devices aim for accuracy by coupling directly to the skull.
- The impact of decoupling on measurement accuracy is not well understood.
Purpose of the Study:
- To investigate how mouthguard decoupling affects kinematic measurement accuracy during simulated sports impacts.
- To quantify the influence of gap size, impact direction, and duration on measurement error.
Main Methods:
- A boil-and-bite mouthguard was fitted to a 3D-printed headform with sensors at the center of gravity.
- Pendulum impact tests were conducted with varying gap sizes (0, 1.6, 4.8 mm) simulating decoupling.
- Linear acceleration and angular velocity were measured and compared to headform reference data.
Main Results:
- Mouthguard measurement error was significantly influenced by gap size, impact duration, and direction.
- Larger gaps and frontal impacts resulted in higher measurement errors.
- Padded conditions also showed increased errors, and some impacts could not be captured.
Conclusions:
- Decoupling significantly impacts the accuracy of boil-and-bite instrumented mouthguards.
- Researchers must be aware of these limitations and consider advanced post-processing techniques.
- Proper fit and understanding decoupling effects are crucial for reliable head kinematics data.
Related Concept Videos
Equipments Used to Measure Body Temperature
2.1K
Body temperature can be assessed using various devices and measured in Celsius or Fahrenheit.
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
2.1K
Equipments Used To Measure Blood Pressure
3.4K
Direct Method
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
3.4K