Related Experiment Video
Updated: Jun 22, 2025

Treating Low Back Pain in Failed Back Surgery Patients with Multicolumn-lead Spinal Cord Stimulation
Published on: June 26, 2018
Biportal Endoscopic Spinal Cord Stimulation Paddle Lead Implantation: Technical Note and Preliminary Clinical Results
Xi-Zi Miao1, Ding-Zhi Gao2, Shao-Mei Yang2
1Department of Pain Medicine, the Second Affiliated Hospital of Chongqing Medical University, Chongqing, China; Department of Spine Surgery, the Second Affiliated Hospital of Chongqing Medical University, Chongqing, China.
Objective:
This study aims to elucidate a novel, minimally invasive surgical technique using a biportal endoscope for the implantation of spinal cord stimulation (SCS) paddle leads and to report the preliminary results of its clinical application.
Materials And Methods:
The perioperative data of patients who underwent the biportal endoscopic SCS paddle lead implantation in our department were collected; the surgical procedure was delineated, and the clinical outcomes were assessed.
Results:
From February 2022 to December 2023, six patients underwent biportal endoscopic SCS paddle lead implantation. The median follow-up time was nine months (range one to three months). The median intraoperative blood loss was 30 mL (range 25-50 mL), and the median operative time was 87.5 minutes (range 75-110 minutes). One patient experienced severe neck pain during the operation, whereas the other five patients experienced no surgical complications. One patient was found to have a slight lead migration three months after surgery, which did not affect the therapeutic effect. The median visual analogue scale (VAS) of the surgical area was 0.5 (range 0-2), 2.5 (range 1-4), and 0.5 (range 0-1) during the operation and one day and one week after the operation, respectively. The median VAS of the six patients' primary disease was 8 (range 7-9) before surgery and 2.5 (range 1-4) at the last postoperative follow-up (pain reduction ≥50%).
Conclusion:
Paddle lead systems for SCS can be implanted successfully using a biportal endoscopic technique.