Related Experiment Video
Updated: Jun 22, 2025

Development of a Lateral Flow Immunochromatographic Strip for Rapid and Quantitative Detection of Small Molecule Compounds
Published on: November 13, 2021
Selection of Single-Analyte Delta Check Rules with Logistic Regression for Detection of Intravenous Fluid
Jianbo Yang1, Sijin Wen2, Christopher R McCudden3
1Department of Pathology, Anatomy and Laboratory Medicine, West Virginia University, Morgantown, WV, United States.
Background:
The conventional single-analyte delta check, utilized for identifying intravenous fluid contamination and other preanalytical errors, is known to flag many specimens reflecting true patient status changes. This study aimed to derive delta check rules that more accurately identify contamination.
Methods:
Results for calcium, creatinine, glucose, sodium, and potassium were retrieved from 326 103 basic or comprehensive metabolic panels tested between February 2021 and January 2022. In total, 7934 specimens showed substantial result changes, of which 1489 were labeled as either contaminated or non-contaminated based on chart review. These labeled specimens were used to derive logistic regression models and to select the most predictive single-analyte delta checks for 4 common contaminants. Their collective performance was evaluated using a test data set from October 2023 comprising 14 717 specimens.
Results:
The most predictive single-analyte delta checks included a calcium change by ≤-24% for both saline and Plasma-Lyte A contamination, a potassium increase by ≥3.0 mmol/L for potassium contamination, and a glucose increase by ≥400 mg/dL (22.2 mmol/L) for dextrose contamination. In the training data sets, multi-analyte logistic regression models performed better than single-analyte delta checks. In the test data set, logistic regression models and single-analyte delta checks demonstrated collective alert rates of 0.58% (95% CI, 0.46%-0.71%) and 0.60% (95% CI, 0.49%-0.74%), respectively, along with collective positive predictive values of 79% (95% CI, 70%-89%) and 77% (95% CI, 68%-87%).
Conclusions:
Single-analyte delta checks selected by logistic regression demonstrated a low false alert rate.
More Related Videos
10:17High-throughput and Comprehensive Drug Surveillance Using Multisegment Injection-Capillary Electrophoresis-Mass Spectrometry
Published on: April 23, 2019
14:42Liquid Chromatography Coupled to Refractive Index or Mass Spectrometric Detection for Metabolite Profiling in Lysate-based Cell-free Systems
Published on: September 23, 2021
Related Concept Videos
Data Validation
Key parameters for method validation include:
High-Performance Liquid Chromatography: Types of Detectors
Difference from Background: Limit of Detection
The LOD indicates the presence or absence...
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...