Related Experiment Video
Updated: Jun 22, 2025

Protocol and Guidelines for Point-of-Care Lung Ultrasound in Diagnosing Neonatal Pulmonary Diseases Based on International Expert Consensus
Published on: March 6, 2019
Comparison of different types of ultrasound probes for lung ultrasound in neonates-A prospective randomized
Lukas Aichhorn1, Lisa Habrina1, Tobias Werther1
1Department of Pediatrics and Adolescent Medicine, Division of Neonatology, Comprehensive Center for Pediatrics, Pediatric Intensive Care & Neuropediatrics, Medical University of Vienna, Vienna, Austria.
Objective:
To determine the effect of different types of probes for lung ultrasound in neonates.
Design:
Prospective, blinded, randomized, comparative study between 2020 and 2022.
Setting:
Single-center study at a third level neonatal unit.
Patients:
Hemodynamically stable infants with either nasal continuous positive airway pressure, high flow nasal cannula or without respiratory support.
Intervention:
Lung ultrasound using either an echo or microconvex probe. As control, the linear probe was used.
Main Outcome Measures:
Primary outcome measure was neonatologist performed lung ultrasound (NPLUS) score. Secondary outcome measures were number of B-Lines, thickness of the pleural line and subjective image quality. Furthermore, correlation between NPLUS results and clinical data was assessed.
Results:
A total of 1584 video loops from 66 patients, with a mean corrected gestational age of 33.8 weeks (SD 4.23) and weight of 1950g (SD 910), respectively, were analyzed. NPLUS score was estimated lower with the echo- and microconvex probe compared to the linear probe, with a coefficient of -2.95 (p < 0.001) and -1.09 (p = 0.19), respectively. Correlation between the pulse oximetric saturation/fraction of inspired oxygen ratio and NPLUS score was moderately strong and best using the microconvex probe (Spearman's rho = -0.63, p<0.001).
Conclusion:
Our results not only confirm the current recommendations, but also demonstrate the extent of the varying results when different probes are used. The differences we discovered call for caution in interpreting scores, especially in the context of guiding therapies and communicating prognoses. Finally, the correlation between NPLUS score and clinical parameters contributes to validating the use of this diagnostic tool.
Related Concept Videos
Ultrasonography
During an ultrasonography procedure, a handheld device called...
Ultrasound I: Abdominal Ultrasonography
Abdominal ultrasonography, commonly known as abdominal ultrasound, is a vital, non-invasive medical imaging technique widely used in healthcare.
Procedure:
This diagnostic tool allows the clinician to visually inspect internal structures within the abdomen, including vital organs such as the liver, gallbladder, pancreas, kidneys, and spleen.
The abdominal ultrasound process begins with applying a special gel to the patient's skin over the abdomen. This gel enhances the...
Ultrasound II: Endoscopic Ultrasound and FibroScan
Endoscopic Ultrasound (EUS):