Halogenated Agent Delivery in Porcine Model of Acute Respiratory Distress Syndrome via an Intensive Care Unit Type Device
Published on: September 24, 2020
Efficiency of passive activated carbon anaesthetic gas capturing systems during simulated ventilation
Christin Wenzel1, Bernd Flamm1, Torsten Loop1
1Department of Anesthesiology and Critical Care, University of Freiburg Medical Center, Faculty of Medicine, University of Freiburg, Freiburg, Germany.
Background:
Interest in passive flow filter systems to remove sevoflurane from anaesthetic machine exhaust have increased recently to mitigate the environmental impact of volatile anaesthetics. These filter systems consist of chemically activated carbon, with limited evidence on their performance characteristics. We hypothesised that their efficiency depends on filter material.
Methods:
Binding capacity was tested for three carbon filter materials (CONTRAfluran®, FlurAbsorb®, and Anaesthetic Agent Filter AAF633). Adsorption efficiency and resistive pressure were determined during simulated ventilation at different stages of filter saturation and fresh gas flow. In addition, sevoflurane concentration in filtered gas was measured at randomly selected anaesthesia workstations.
Results:
Sevoflurane concentration in filtered gas exceeded 10 ppm when saturated with 184 ml sevoflurane each for CONTRAfluran and FlurAbsorb and 276 ml for AAF633. During simulated ventilation, sevoflurane concentration >10 ppm passed through CONTRAfluran and AAF633 at fresh gas flow 10 L min-1 only at maximum saturation, but through FlurAbsorb at all stages of saturation. The resistance pressure of all filters was negligible during simulated ventilation, but increased up to 5.2 (0.2) cm H2O during simulated coughing. At two of seven anaesthesia workstations, sevoflurane concentration in filtered exhaust gas was >10 ppm.
Conclusions:
Depending on the filter material and saturation, the likelihood of sevoflurane passing through passive flow carbon filters depends on the filter material and fresh gas flow. Combining the filter systems with anaesthetic gas scavenging systems could protect from pollution of ambient air with sevoflurane.
More Related Videos
Related Concept Videos
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Acute Respiratory Failure-V
Ensure that patients are monitored continuously for their response to therapy, including changes in...
Acute Respiratory Failure-III
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...