Effects of Faraday cup deterioration on Sr and Cr isotope analyses by thermal ionization mass spectrometry
Jonas M Schneider1,2, Thorsten Kleine1,2
1Max Planck Institute for Solar System Research Justus-von-Liebig-Weg 3 D-37077 Göttingen Germany schneiderj@mps.mpg.de +49 551 384979 308.
Abstract:
By comparing data from an extensive set of Sr and Cr isotope measurements performed on two different thermal ionization mass spectrometers (TIMS), using three sets of Faraday cups with different usage histories, we assess the effects of Faraday cup deterioration on high-precision isotope measurements by TIMS. We find that dynamic 84Sr/86Sr and 87Sr/86Sr measurements provide stable and reproducible results over the entire 56 months of this study, regardless of which set of Faraday cups is used. By contrast, static 84Sr/86Sr and 87Sr/86Sr measurements lead to deviant results, drifts over time, and in general exhibit larger scatter. For the most part, these differences can be attributed to changing Faraday cup efficiencies. For the instruments of this study we find that the center cup is most affected, consistent with this cup often receiving the highest ion beam intensities during measurements conducted in our laboratory. For Cr isotopes, we find that the correlation between mass fractionation-corrected 53Cr/52Cr and 54Cr/52Cr ratios observed for static measurements in several prior studies also reflects different Faraday cup efficiencies. Again, the changing efficiency of predominantly the center cup can account for the observed drift and correlation in 53Cr/52Cr and 54Cr/52Cr. Multi-static Cr isotope measurements reduce this drift, but still result in a residual correlation between the two ratios, suggesting this correlation in part also reflects unaccounted mass fractionation effects.
More Related Videos
08:43Separation of Uranium and Thorium for 230Th-U Dating of Submarine Hydrothermal Sulfides
Published on: May 20, 2019
12:18Co-localizing Kelvin Probe Force Microscopy with Other Microscopies and Spectroscopies: Selected Applications in Corrosion Characterization of Alloys
Published on: June 27, 2022
Related Concept Videos
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
Qualitative Analysis
For instance, group IV...
Atomic Spectroscopy: Effects of Temperature
At thermal equilibrium, the relative populations of excited and ground state atoms can be estimated using the Maxwell–Boltzmann distribution. For example, an increase in temperature...
Atomic Mass
Mass Spectrometry: Isotope Effect
Atomic Absorption Spectroscopy: Interference
Spectral interference occurs when signals from other elements or molecules overlap with the analyte signal, falsely elevating or masking the analyte's absorbance. This interference can be corrected using Zeeman,...