Aveir VR, retrievable leadless pacing in the young.
Ashley Wong1, Jay Yeh1, Stacy Davidson1
1Department of Pediatrics and Pediatric Cardiology, University of California at Davis, Sacramento, California, USA.
Pacing and Clinical Electrophysiology : PACE
|July 5, 2024
Summary
Leadless pacemaker implantation using the Aveir VR system is safe and feasible in young patients. This study demonstrates stable pacing parameters and no complications in pediatric patients, supporting its use in this population.
Area of Science:
- Cardiology
- Pediatric Electrophysiology
- Medical Device Technology
Background:
- Successful implantation of the Aveir VR leadless pacemaker has been documented in adults.
- Limited data exists on the safety and feasibility of Aveir VR implantation in pediatric populations.
Purpose of the Study:
- To evaluate the safety and feasibility of Aveir VR leadless pacemaker implantation in patients aged 21 years or younger.
- To assess pacing parameters and outcomes in young patients following Aveir VR implantation.
Main Methods:
- A retrospective observational study was conducted on 10 pediatric patients (≦21 years) who underwent Aveir VR implantation.
- Implantation was performed via internal jugular or femoral venous approaches.
- Data collected included patient demographics, pacing thresholds, longevity, and complications at implantation and follow-up.
Main Results:
- Ten pediatric patients (median age 12.5-17 years, 80% male) received Aveir VR leadless pacemakers.
- Implantations were primarily via the internal jugular vein (90%).
- Indications included heart block (60%) and sinus pauses (40%). Pacing parameters remained adequate through a median 9-month follow-up, with a predicted longevity of 23.8 years and no reported complications.
Conclusions:
- Aveir VR leadless pacemaker implantation is a feasible and safe procedure in the pediatric population.
- Stable pacing parameters and good predicted longevity support the use of Aveir VR in young patients.
- The internal jugular and femoral venous approaches are viable for leadless pacemaker implantation in this demographic.