The impact of the Impeller's hub design on the performance and blood damage in a microaxial mechanical circulatory
Yuval Gabso1, Moshe Rosenfeld2, Idit Avrahami3
1School of Mechanical Engineering, Tel Aviv University, Tel Aviv, 69978, Israel; Department of Mechanical Engineering and Mechatronics, Ariel University, Israel.
Abstract:
This study uses CFD methods to investigate the effects of the impeller's geometry on the hemodynamic characteristics, pump performance, and blood damage parameters, in a percutaneous microaxial Mechanical Circulatory Support (MCS) device. The numerical simulations employ the steady state Reynolds-Averaged Navier-Stokes approximation using the SST k-ω turbulent model. Three different impeller models are examined with different hub conversion angles (α = 0○, 3○ and 5○). The analysis includes 23 cases for different pressure heads (Δp = 60-80 mmHg) and angular velocities (ω = 30-52 kRPM). The obtained flow rate is compared between the cases to assess the effect of the impeller's design and working conditions on the pump performance. The comparative risk of shear-induced platelet activation is estimated using the statistical median of the stress-accumulation values calculated along streamlines. The risk of hemolysis is estimated using the average exposure time to shear stress above a threshold (τ > 425 Pa). The results reveal that the shape of the impeller's hub has a great impact on the flow patterns, performance, and risk of blood damage, as well as the angular velocity. The highest flow rate (Q = 3.7 L/min) and efficiency (η = 11.3 %) were achieved using a straight hub (α = 0○). Similarly, for the same condition of flow and pressure, the straight hub impeller has the lowest blood damage risk parameters. This study shed light on the effect of pump design on the performance and risk of blood damage, indicating the roles of the hub shape and angular velocity as dominant parameters.
Related Concept Videos
Design of Transmission Shafts
Design of Transmission Shafts - Stress Analysis
Transmission Shafts: Problem Solving
Next, use bending moment diagrams for the shaft to...
Design Example: Deciding Thickness of Lubricating Fluid in a Shaft
To calculate the required thickness of the lubricant layer, the tangential velocity at the shaft's surface must first be determined. This velocity is calculated by converting the rotational speed to angular velocity...