Dynamic Ultrasound Localization Microscopy Without ECG-Gating.
Nin Ghigo1, Gerardo Ramos-Palacios2, Chloé Bourquin1
1Department of Engineering Physics, Polytechnique Montréal, Montréal, Quebec, Canada.
Ultrasound in Medicine & Biology
|July 5, 2024
Summary
This study introduces a motion-matching technique for Dynamic Ultrasound Localization Microscopy (DULM), eliminating the need for ECG-gating. This advancement simplifies DULM experiments and enhances accessibility for microbubble-based flow velocity measurements.
Area of Science:
- Biomedical Engineering
- Medical Imaging
- Ultrasound Technology
Background:
- Dynamic Ultrasound Localization Microscopy (DULM) enables non-invasive pulsatility measurements in the rodent brain by tracking microbubbles (MBs).
- Traditional DULM methods require ECG-gating, which limits applications and increases acquisition time.
- There is a need for DULM techniques that do not rely on external synchronization signals.
Purpose of the Study:
- To eliminate the requirement for ECG-gating in Dynamic Ultrasound Localization Microscopy (DULM) experiments.
- To introduce a novel motion-matching method for accurate time registration in DULM.
- To enable DULM on a wider range of datasets without compromising data quality.
Main Methods:
- Developed a motion-matching algorithm utilizing tissue Doppler signals to leverage inherent cyclic brain tissue motion.
- Estimated tissue Doppler for frame groups at locations of local maxima on the skull's surface.
- Time-registered frame groups by aligning their tissue Doppler signals to a reference group, synchronizing with the cardiac cycle.
Main Results:
- The motion-matching method was successfully validated in rat, mice (2D), and feline (3D) models.
- Results from motion-matching time registration were comparable to those obtained using ECG-gating.
- Demonstrated the extraction of dynamic velocity and density cine-loops and estimation of pulsatility index without ECG data.
Conclusions:
- DULM can be performed effectively without external gating, broadening its applicability to any ultrasound localization microscopy dataset with sufficient microbubble detection.
- Motion-matching for time registration significantly advances DULM by simplifying experimental procedures and increasing accessibility.
- This technique enhances the utility of DULM for non-invasive hemodynamic assessments in various animal models.