Evaluation of low-dose pediatric chest CT examination using in-house developed various age-size pediatric chest

Dafa Miftahuddin1, Audiena Gelung Prayitno1, Aditya Prayugo Hariyanto1

  • 1Department of Physics, Institut Teknologi Sepuluh Nopember, Kampus ITS - Sukolilo Surabaya 600111, East Java, Indonesia.

PubMed

Insights

Developing pediatric chest phantoms (VAPC) revealed that smaller children receive higher radiation doses during low-dose CT scans. Optimizing protocols with low kV and iterative reconstruction improves image quality and noise reduction for pediatric patients.

Area of Science:

  • Medical Imaging Physics
  • Pediatric Radiology
  • Radiation Dosimetry

Background:

  • Pediatric patients present unique challenges in CT imaging due to significant variations in size and age.
  • Optimizing low-dose CT protocols is crucial to minimize radiation exposure while maintaining diagnostic image quality in children.

Purpose of the Study:

  • To develop Various Age-size Pediatric Chest Phantoms (VAPC) for evaluating low-dose CT protocols.
  • To assess organ-specific doses and image quality parameters under different low-dose CT conditions in pediatric phantoms.

Main Methods:

  • VAPC phantoms were created based on pediatric data from 1, 4, and 7-year-olds.
  • Six CT protocols (standard dose, low mA, low kV) using Filtered Back Projection (FBP) and iterative reconstruction (IR) were evaluated.
  • Organ doses (lungs, heart, spinal cord) were measured, and image quality metrics (linearity, MTF, CNR, NPS) were assessed.

Main Results:

  • Mean organ-specific doses exceeded CTDIvol, with younger phantoms receiving significantly higher lung doses.
  • Low kV protocols resulted in lower organ doses compared to low mA protocols.
  • Age significantly impacted organ dose, MTF, CNR, and NPS, while CT number linearity remained unbiased at low doses.

Conclusions:

  • Smaller pediatric patients are exposed to higher doses in low-dose CT, while larger children exhibit reduced contrast resolution and increased noise.
  • Low kV combined with FBP enhanced spatial resolution.
  • Low mA with IR improved noise reduction for better detection of low-contrast objects.
Abstract