Area under the curve and amplitude of the compound motor action potential are clinically interchangeable quantitative
Richard H Epstein1, Réka Nemes2, Johnathan R Renew3
1Department of Anesthesiology, Perioperative Medicine & Pain Management, University of Miami, Miller School of Medicine, Miami, FL, USA.
Background:
Current guidelines recommend quantitative neuromuscular block monitoring during neuromuscular blocking agent administration. Monitors using surface electromyography (EMG) determine compound motor action potential (cMAP) amplitude or area under the curve (AUC). Rigorous evaluation of the interchangeability of these methods is lacking but necessary for clinical and research assurance that EMG interpretations of the depth of neuromuscular block are not affected by the methodology.
Methods:
Digitised EMG waveforms were studied from 48 patients given rocuronium during two published studies. The EMG amplitudes and AUCs were calculated pairwise from all cMAPs classified as valid by visual inspection. Ratios of the first twitch (T1) to the control T1 before administration of rocuronium (T1c) and train-of-four ratios (TOFRs) were compared using repeated measures Bland-Altman analysis.
Results:
Among the 2419 paired T1/T1c differences where the average T1/T1c was ≤0.2, eight (0.33%) were outside prespecified clinical limits of agreement (-0.148 to 0.164). Among the 1781 paired TOFR differences where the average TOFR was ≥0.8, 70 (3.93%) were outside the prespecified clinical limits of agreement ((-0.109 to 0.134). Among all 7286 T1/T1c paired differences, the mean bias was 0.32 (95% confidence interval 0.202-0.043), and among all 5559 paired TOFR differences, the mean bias was 0.011 (95% confidence interval 0.0050-0.017). Among paired T1/T1c and TOFR differences, Lin's concordance correlation coefficients were 0.98 and 0.995, respectively. Repeatability coefficients for T1/T1c and TOFR were <0.08, with no differences between methods.
Conclusions:
Quantitative assessment neuromuscular block depth is clinically interchangeable when calculated using cMAP amplitude or the AUC.
Related Concept Videos
Nondepolarizing (Competitive) Neuromuscular Blockers: Pharmacokinetics
Instead, they are transported by the blood to different tissues. Muscles with a greater blood supply (arteries) and blood flow receive more...
Nondepolarizing (Competitive) Neuromuscular Blockers: Mechanism of Action
Competitive antagonists prevent acetylcholine from binding to its receptor, inhibiting membrane depolarization. Without conformational changes or intrinsic...
Neuromuscular Junction And Blockade
Nondepolarizing (Competitive) Neuromuscular Blockers: Pharmacological Actions
Although all competitive neuromuscular blockers are designed...
Depolarizing Blockers: Mechanism of Action
Succinylcholine is the most commonly used depolarizing blocker. Chemically, it constitutes two molecules of acetylcholine joined together by an acetate methyl group. They act on the receptors in the same way as acetylcholine. Because...
Depolarizing Blockers: Pharmocokinetics