Preventing Iatrogenic Fibula Fractures Using the Push-Pull Technique: A Biomechanical Comparison of Unicortical
Background:
Displaced diaphyseal fractures can be reduced using the push-pull technique, wherein a plate is affixed to the distal fragment of the fracture, a post screw is placed proximal to the plate, and a lamina spreader creates distraction. This study evaluated the load to failure and mechanism of failure of bicortical and unicortical post screws during reduction.
Materials And Methods:
Four matched pairs of cadaver legs were subjected to a 2-cm oblique osteotomy simulating a displaced, oblique diaphyseal fracture. A 6-hole compression plate was affixed to the distal fragment with 2 unicortical locking screws, and a 12-mm unicortical or 20-mm bicortical screw was inserted as a post screw proximal to the plate. A lamina bone spreader was used to exert a distraction force between the plate and the post screw. A mechanical actuator simulated the distraction procedure until failure. Maximum applied load, displacement, and absorbed energy were recorded and compared across unicortical and bicortical groups by paired t tests.
Results:
At maximum load, we found statistically significant differences in displacement (P=.003) and energy absorbed (P=.022) between the two groups. All unicortical screws failed through screw toggle and bone cut-out. Bicortical screws failed through bending, with no visible damage to the bone at the screw site.
Conclusion:
When diaphyseal fractures are significantly shortened and require a greater distraction force to achieve reduction, bicortical screws demonstrate a higher mechanical load to failure and increased bone loss from the screw-removal site. A unicortical post screw may be used if minimal distraction is needed. [Orthopedics. 2024;47(5):308-312.].