Aerosol Delivery Efficiency With High-Flow Nasal Cannula Therapy in Neonatal, Pediatric, and Adult Nasal Upper-Airway
Robert M DiBlasi1, Rebecca J Engberg2, Jonathan Poli3
1Respiratory Therapy Department, Seattle Children's Hospital, Seattle, Washington; and Center for Respiratory Biology and Therapeutics, Seattle Children's Research Institute, Seattle, Washington robert.diblasi@seattlechildrens.org.
Background:
High-flow nasal cannula (HFNC) systems employ different methods to provide aerosol to patients. This study compared delivery efficiency, particle size, and regional deposition of aerosolized bronchodilators during HFNC in neonatal, pediatric, and adult upper-airway and lung models between a proximal aerosol adapter and distal aerosol circuit chamber.
Methods:
A filter was connected to the upper airway to a spontaneously breathing lung model. Albuterol was nebulized using the aerosol adapter and circuit at different clinical flow settings. The aerosol mass deposited in the upper airway and lung was quantified. Particle size was measured with a laser diffractometer. Regional deposition was assessed with a gamma camera at each nebulizer location and patient model with minimum flow settings.
Results:
Inhaled lung doses ranged from 0.2-0.8% for neonates, 0.2-2.2% for the small child, and 0.5-5.2% for the adult models. Neonatal inhaled lung doses were not different between the aerosol circuit and adapter, but the aerosol circuit showed marginally greater lung doses in the pediatric and adult patient models. Impacted aerosols and condensation in the non-heated HFNC and aerosol delivery components contributed to the dispersion of coarse liquid droplets, high deposition (11-44%), and occlusion of the supine neonatal upper airway. In contrast, the upright pediatric and adult upper-airway models had minimal deposition (0.3-7.0%) and high fugitive losses (∼24%) from liquid droplets leaking out of the nose. The high impactive losses in the aerosol adapter (56%) were better contained than in the aerosol circuit, resulting in less cannula sputter (5% vs 22%), fewer fugitive losses (18% vs 24%), and smaller inhaled aerosols (5 µm vs 13 µm).
Conclusions:
The inhaled lung dose was low (1-5%) during HFNC. Approaches that streamline aerosol delivery are needed to provide safe and effective therapy to patients receiving aerosolized medications with this HFNC system.
More Related Videos
15:04Modeling and Simulations of Olfactory Drug Delivery with Passive and Active Controls of Nasally Inhaled Pharmaceutical Aerosols
Published on: May 20, 2016
07:28Dry Powder and Nebulized Aerosol Inhalation of Pharmaceuticals Delivered to Mice Using a Nose-only Exposure System
Published on: April 6, 2017
Related Concept Videos
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Administering Oxygen by Nasal Cannula
Nasal Cannulas
A nasal cannula is a lightweight tube split into two prongs placed in the nostrils,...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Suctioning the Nasopharyngeal Airway
Equipment Required
Inhaled Medications