Related Experiment Video
Updated: Jun 21, 2025

Author Spotlight: Improving Radiation Therapy Access with Radiation Planning Assistant
Published on: October 6, 2023
A method for empirically validating FMEA RPN scores in a radiation oncology clinic using physics QC data
Leonard H Kim1, Badal R Juneja1, Natalie N Viscariello1
1Department of Radiation Oncology, MD Anderson Cancer Center at Cooper, Camden, New Jersey, USA.
Abstract:
In failure modes and effects analysis (FMEA), the components of the risk priority number (RPN) for a failure mode (FM) are often chosen by consensus. We describe an empirical method for estimating the occurrence (O) and detectability (D) components of a RPN. The method requires for a given FM that its associated quality control measure be performed twice as is the case when a FM is checked for in an initial physics check and again during a weekly physics check. If instances of the FM caught by these checks are recorded, O and D can be computed. Incorporation of the remaining RPN component, Severity, is discussed. This method can be used as part of quality management design ahead of an anticipated FMEA or afterwards to validate consensus values.

