Efficacy of estimations of Hartmann-Shack sensors in small pupil sizes
Francisco de Asís Bartol-Puyal1,2,3, Galadriel Giménez1,2, Silvia Méndez-Martínez2
1Department of Ophthalmology, Miguel Servet University Hospital, Zaragoza, Spain.
Purpose:
The purpose of this study was to evaluate the in vivo efficacy of the estimations of wavefront analyzers using Hartmann-Shack technology to measure optical aberrations when the pupil size is smaller than the evaluated pupil area.
Methods:
Patients implanted with the monofocal ZCB00 intraocular lens (Johnson and Johnson) were examined with the KR-1W Wavefront Analyzer (Topcon) without pharmacological mydriasis and with it afterward. Optical aberrations were analyzed considering a 4-mm pupil and a 6-mm pupil for both examinations.
Results:
Sixty-six eyes of 33 patients with a mean axial length of 23.35 ± 0.91 mm were assessed. The mean pupil diameter at the baseline examination was 5.05 ± 0.88 mm and under pharmacological mydriasis, it was 6.29 ± 0.84 mm. Outcomes were similar with and without dilation in the 4-mm comparison. However, there was a great disparity in the 6-mm comparison. Most of the values obtained under mydriasis were statistically lower than at baseline (P < 0.05).
Conclusion:
The iris interferes with measurements of wavefront aberrations, and therefore, real pupil size should always be checked before evaluating optical aberrations with Hartman-Shack sensors. When pupil size is smaller than the analyzed diameter, ocular, and internal, and sometimes, corneal aberrations are estimated far more positive than real values.