Cardiac Cine MRI Using a Commercially Available 0.55-T Scanner
Martin Segeroth1, David J Winkel1, Jan Vosshenrich1
1From the Department of Radiology (M.S., D.J.W., J.V., H.C.B., J.B., F.S., M.P.) and Clinic of Cardiology (P.H., M.J.Z.), University Hospital Basel, Petersgraben 4, 4031 Basel, Switzerland; and Magnetic Resonance, Siemens Healthcare, Erlangen, Germany (D.G.).
Low-field 0.55-T cardiac cine MRI provides comparable biventricular volume and function results to 1.5-T scanners. This comparison study confirms 0.55-T MRI as a viable alternative for cardiovascular MRI, especially when acquisition time is adjusted.
Area of Science:
- Cardiovascular Imaging
- Magnetic Resonance Imaging (MRI)
- Cardiac Function Assessment
Background:
- Standard cardiac cine MRI utilizes 1.5-T scanners for assessing left ventricular (LV) and right ventricular (RV) function.
- Low-field strength MRI scanners (0.55 T) offer potential advantages in accessibility and cost but require validation for quantitative cardiac analysis.
Purpose of the Study:
- To compare quantitative biventricular volumetric and functional parameters derived from a 0.55-T low-field cardiac cine MRI scanner against a standard 1.5-T scanner.
- To evaluate the diagnostic equivalence of 0.55-T cardiac MRI for clinical cardiovascular assessments.
Main Methods:
- Prospective study involving healthy volunteers and patients undergoing same-day cine imaging on both 0.55-T and 1.5-T MRI scanners.
- Volumetric and functional parameters were assessed by two independent expert readers.
- Data analysis included Bland-Altman plots to visualize agreement and calculate mean correlation and deviation between the two systems.
Main Results:
- Very strong mean correlation (r = 0.98) was observed between the 0.55-T and 1.5-T systems for both healthy volunteers and patients.
- Average deviation in LV and RV parameters was minimal, reported as 1.6% in volunteers and 1.1% in patients, indicating high agreement.
- Comparable quantitative biventricular volumetric and functional parameters were achieved with the 0.55-T scanner, provided acquisition time was doubled.
Conclusions:
- Cardiac cine MRI performed on a 0.55-T low-field scanner yields results comparable to those obtained with a 1.5-T scanner for biventricular parameters.
- The 0.55-T system demonstrates high reliability and agreement for quantitative assessment of cardiac function and volume.
- 0.55-T cardiac MRI is a viable alternative to 1.5-T MRI for cardiovascular imaging, with careful consideration of acquisition protocols.
More Related Videos
11:13Quantification of Mouse Heart Left Ventricular Function, Myocardial Strain, and Hemodynamic Forces by Cardiovascular Magnetic Resonance Imaging
Published on: May 24, 2021
12:24Noninvasive Assessment of Cardiac Abnormalities in Experimental Autoimmune Myocarditis by Magnetic Resonance Microscopy Imaging in the Mouse
Published on: June 20, 2014
Related Concept Videos
Magnetic Resonance Imaging
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...