Local impedance drop-guided versus lesion size index-guided pulmonary vein isolation
Evgeny Lian1, Robert Pantlik2, Vera Maslova3
1Department of Internal Medicine III (Cardiology and Intensive Care Medicine), University Hospital Schleswig-Holstein (UKSH), Kiel, Germany. lyanevgeny@gmail.com.
Background:
Local tissue impedance drop (LID) and lesion size index (LSI) technologies are valuable for predicting effective lesion formation. This study compares the acute and long-term efficacy of LID-guided versus LSI-guided pulmonary vein isolation (PVI) for atrial fibrillation treatment.
Methods:
We retrospectively analyzed two patient groups undergoing radiofrequency PVI. In the LID-guided group (n = 35), ablation was performed without contact force monitoring, stopping at the LID plateau (target LID 12 Ohm posterior, 16 Ohm anterior). In the LSI-guided group (n = 31), ablation used contact force information with target LSI (5 anterior, 4 posterior). Both groups utilized a power of 40 W anterior and 30 W posterior, with < 6 mm inter-lesion distance. Gap mapping and touch-up ablation were done if necessary.
Results:
PVI was achieved with a significantly shorter ablation time in the LSI-guided group (25 min [21;31] vs 30 [27;35], p = 0.035). PV gaps were more frequent in the LID-guided group (74% vs 42%, p = 0.016). Over 11.5 ± 2.9 months follow-up, arrhythmia recurrence was higher in the LID-guided group (34.3% vs 16.1%, p = 0.037). A redo procedure performed in 10 (28.6%) patients in the LID-guided group and 3 (9.7%) in the LSI-guided group showed chronic PV reconnections in 7 out of 10 (70%) and 2 out of 3 (67%) patients, respectively.
Conclusions:
LSI-guided ablation results in shorter ablation time and fewer PV gaps compared to LID-guided ablation. Despite initial success, LID-guided ablation had higher arrhythmia recurrence and PV reconnections during long-term follow-up compared to LSI-guided ablation.
More Related Videos
06:48Three-Dimensional Echocardiographic Method for the Visualization and Assessment of Specific Parameters of the Pulmonary Veins
Published on: October 28, 2020
05:43Author Spotlight: Improving Lesion Contiguity in Pulmonary Vein Isolation via Proactive Esophageal Cooling
Published on: April 19, 2024
Related Concept Videos
Acute Respiratory Failure-II
The underlying physiological abnormalities that contribute to hypoxemic respiratory failure include:
Mitral Stenosis I: Introduction
Atelectasis II: Pathophysiology