Wearable cuffless blood pressure tracking: when will they be good enough?
Aletta E Schutte1,2,3,4
1School of Population Health, University of New South Wales, Sydney, NSW, Australia. a.schutte@unsw.edu.au.
Cuffless blood pressure monitoring devices show potential but require further validation. A 24-hour self-test highlighted challenges, suggesting current clinical practices for blood pressure measurement need reevaluation.
Area of Science:
- Biomedical Engineering
- Cardiovascular Health
- Wearable Technology
Background:
- Wearable health monitoring is a significant industry, with cuffless blood pressure monitoring being a key goal.
- Raised blood pressure is a global health concern and a leading cause of mortality.
- Despite numerous attempts and market entries, cuffless blood pressure devices face challenges in clinical acceptance.
Purpose of the Study:
- To evaluate the challenges and opportunities in cuffless blood pressure monitoring.
- To compare cuffless devices against traditional cuff-based measurements during a 24-hour period.
- To assess the feasibility of cuffless devices for clinical decision-making.
Main Methods:
- A 24-hour self-monitoring blood pressure test was conducted.
- Five devices were used simultaneously: two validated cuff-based devices (left and right arm) and three cuffless wearable devices (wrist-band, chest patch, ring).
- Measurements were taken during rest, sleep, and a high-stress event (Rugby World Cup final).
Main Results:
- None of the tested cuffless devices perfectly captured blood pressure during all physiological conditions.
- The study highlighted the variability and challenges in obtaining accurate blood pressure readings, especially during dynamic physiological states.
- The limitations observed question the adequacy of single blood pressure measurements in clinical practice.
Conclusions:
- Current cuffless blood pressure monitoring technologies face significant challenges in accuracy and reliability.
- The study underscores the need for revisiting current clinical practices for blood pressure assessment.
- Further research and technological advancements are required before cuffless measurements can be reliably integrated into clinical decision-making.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Measurement of Blood Pressure
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Errors occurring during blood pressure monitoring
Several factors...
Assessment of blood pressure in brachial artery(two-step method)