Methane Quantification Performance of the Quantitative Optical Gas Imaging (QOGI) System Using Single-Blind
Chiemezie Ilonze1, Jiayang Lyra Wang2,3, Arvind P Ravikumar2,3
1Department of Mechanical Engineering, Colorado State University, Fort Collins, CO 80523, USA.
Sensors (Basel, Switzerland)
|July 13, 2024
Summary
This study evaluated the FLIR QL320 quantitative optical gas imaging (QOGI) system for natural gas leak detection. The system showed variable accuracy, with most measurements within a factor of 3, improving under optimal conditions.
Area of Science:
- Environmental Science
- Chemical Engineering
- Instrumentation & Measurement
Background:
- Optical gas imaging (OGI) is crucial for detecting methane (CH4) leaks in the oil and gas industry.
- Quantitative OGI (QOGI) systems offer potential for rapid, accurate leak quantification.
- Standardized evaluation of QOGI performance is necessary for widespread adoption.
Purpose of the Study:
- To assess the quantification accuracy of the FLIR QL320 QOGI system.
- To evaluate system performance under realistic, near-field conditions mimicking oil and gas operations.
- To identify factors influencing QOGI measurement accuracy.
Main Methods:
- Single-blind experiments were conducted at an outdoor, controlled testing facility.
- 357 measurements were taken for controlled releases of compressed natural gas (0.1–2.9 kg CH4/h).
- Data were analyzed across 26 release scenarios and 71 camera positions.
Main Results:
- 75% of individual measurements fell within a quantification factor of 3 (error: -67% to 200%).
- Individual measurement errors ranged from -90% to 831%.
- Averaging estimates from multiple camera positions reduced errors to -79% to +297%.
Conclusions:
- QOGI system performance is dependent on environmental and operational factors.
- Accuracy improved with higher release rates, clear skies, and low wind speeds (≤1 mph).
- The FLIR QL320 shows promise but requires careful consideration of conditions for reliable methane leak quantification.