Related Experiment Video
Updated: Jun 21, 2025

Author Spotlight: Advancing Endoscopic Ossiculoplasty – Techniques, Innovations, and Practical Guidance for Clinical Integration
Published on: January 26, 2024
Safety and performance of the new med-el total ossicular replacement prostheses
Esther Schimanski1, Thomas Lenarz2, Susan Busch2
1ENT Practice, Zentrum fuer Mittelohrchirurgie (Centre for Middle Ear Surgery), 44536, Luenen, Germany. e.schimanski@mittelohr.de.
Purpose:
This multicentric, retrospective study provides safety and performance data of the MED-EL total ossicular replacement prostheses (TORP).
Methods:
Patients underwent tympanoplasty with mXACT Total Prosthesis Center, mXACT Total Prosthesis Offcenter or mXACT PRO Total Prosthesis. The clinical data were retrospectively analyzed. Follow-up examination included access to the medical record (for adverse events) of the patients, ear microscopy and pure-tone audiometry to determine the post-operative pure tone average of the frequencies 0.5, 1, 2 and 3 kHz (PTA4). The post-operative PTA4 air bone gap (ABG) was used to evaluate the audiological outcome. A post-operative minimum and maximum follow-up period was not defined.
Results:
103 patients were implanted with a TORP. 102 (88 adults, 14 children; 37 CHL, 64 MHL, 1 not specified) patients were analyzed for safety and 92 (79 adults, 13 children) patients for performance of the prostheses. ADVERSE EVENTS RESULTS (N = 102): In 1 patient (child, mXACT Total Prosthesis Offcenter) a prosthesis dislocation was reported, which lead to a revision surgery. No prosthesis extrusion or migration was reported. AUDIOLOGICAL RESULTS (N = 92): 49 (53.3%) of the 92 patients had a PTA4 ABG of ≤ 20 dB and therefore a successful rehabilitation. The mean post-operative PTA4 ABG of all 92 patients was 21.0 ± 9.7 dB. The first endpoint, improvement in post-operative PTA4 ABG of ≤ 20 dB by ≥ 25% of the patients was achieved. The individual Δ BC PTA4 (post-operative minus pre-operative BC PTA4) thresholds were stable (within ± 5 dB HL) in 91 (98.9%) patients. 1 patient had a BC PTA4 deterioration of 11.3 dB HL.
Conclusion:
The MED-EL TORPs are safe and effective for middle ear reconstruction. Trial registration number NCT05565339, September 09, 2022, retrospectively registered.