A novel method for highly linear gradient coil design based on stream function
Related Experiment Video
Updated: Jun 21, 2025

MRM Microcoil Performance Calibration and Usage Demonstrated on Medicago truncatula Roots at 22 T
Published on: January 16, 2021
Yufu Zhou1, Zhengrong Liu1, Qing Zhang1
1Medical Imaging Center, Department of Electronic Engineering and Information Science, University of Science and Technology of China, Hefei, China.
Background:
Magnetoencephalography (MEG) and magnetic resonance imaging (MRI) are non-invasive imaging techniques that offer effective means for disease diagnosis. A more straightforward and optimized method is presented for designing gradient coils which are pivotal parts of the above imaging systems.
Purpose:
A novel design method based on stream function combining an optimization algorithm is proposed to obtain highly linear gradient coil.
Methods:
Two-dimensional Fourier expansion of the current field on the surface where the coil is located and the equipotential line of the expansion term superposition according to the number of turns of the coil are used to represent the coil shape. Particle swarm optimization is utilized to optimize the coil shape while linearity and field uniformity are used as parameters to evaluate the coil performance. Through this method, the main parameters such as input current distribution region, coil turns, desired magnetic field strength, expansion order and iteration times can be combined in a given solution space to optimize coil design.
Results:
Simulation results show that the maximum linearity spatial deviation of the designed bi-planar x-gradient coil compared with that of target field method is reduced from 14% to 0.54%, and that of the bi-planar z-gradient coil is reduced from 8.98% to 0.52%. Similarly, that of the cylindrical x-gradient coil is reduced from 2% to 0.1%, and that of the cylindrical z-gradient coil is reduced from 0.87% to 0.45%. The similar results are found in the index of inhomogeneity error. Moreover, it has also been verified experimentally that the result of measured magnetic field is consist with simulated result.
Conclusions:
The proposed method provides a straightforward way that simplifies the design process and improves the linearity of designed gradient coil, which could be beneficial to realize better magnetic field in engineering applications.
Related Concept Videos
Magnetic Field Of A Current Loop
Magnetic Field Due To A Thin Straight Wire
Magnetic Field Due to Two Straight Wires
Bernoulli's Equation for Flow Along a Streamline
Magnetic Field of a Solenoid
Consider a solenoid with 100 turns wrapped around a cylinder of...
Bernoulli's Equation for Flow Normal to a Streamline
The pressure difference depends on the fluid's velocity and radius of curvature. The pressure variation is minimal in flows with nearly straight streamlines.