Accuracy of point-of-care capillary blood sugar measurements in critically ill patients: An observational study
Keshabanand Mishra1, Shivanand Mishra1, Tanmay Katial1
1Department of Anaesthesiology, Vardhman Mahavir Medical College and Safdarjung Hospital, New Delhi, India.
Background:
Accurately monitoring blood glucose levels is vital for critically ill individuals. Point-of-care (POC) glucose meters are commonly used in local intensive care units (ICUs). This study aimed to assess the precision of POC glucose meter readings in critically ill individuals with specific evaluation in patients with and without shock against the reference standard of venous blood glucose measurements.
Methods:
An observational study was done on adult patients admitted in the ICU at a teaching institution. Capillary blood samples were collected from the patient's fingertip using lancet device with aseptic measures. The sample was analyzed using the GlucoCare Sense Glucometer (RMD Mediaids Limited, Taiwan). At the same time, 2 ml of blood was drawn from the patient's peripheral veins and analyzed by glucose oxidase-peroxidase method as reference.
Results:
POC glucose measurements averaged 140 ± 20.23 mg/dl, while laboratory values were recorded as 116.10 ± 17.13 mg/dl. The difference between the two methods was 24.34 ± 12.01 mg/dl. A strong correlation (r = 0.805) was found between capillary and laboratory blood glucose levels, indicating a significant association (P < 0.0001). Twenty-two (44%) patients were in shock during the study. The mean difference between laboratory and POC blood glucose levels was higher in patients with circulatory shock (36.82 ± 4.84 mg/dl) than those without shock (14.61 ± 4.49 mg/dl), P < 0.05.
Conclusion:
POC glucose meters may lead to underdetection of hypoglycemia in critically ill patients, as their values are higher than laboratory values. Moreover, the results showed that POC glucometers are inaccurate for monitoring glucose in hypotensive patients in shock. Standard venous glucose monitoring methods may be more appropriate for these patients.
Related Concept Videos
Errors occurring during blood pressure monitoring
Several factors...
SBAR II: Application of SBAR
SBAR Report from a Nurse to a Health Care Provider
S: "Hello, Dr. Smith. This is Jane, RN, from the Med Surg unit. I am calling to tell you about Ms. White in Room 210, who is experiencing increased pain and redness at her incision site. Her recent...
Hypoglycemia and Glucagon
Diabetes: Symptoms, Diagnosis, and Complications
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Pre-Procedural Guidelines for Assessing Blood Pressure