A Spine Robotic-Assisted Navigation System for Pedicle Screw Placement
Published on: May 11, 2020
An Integrated System Using Ultrasound-Based Registration for Surgical Navigation of Iliosacral Screw Insertions
Objective:
Pelvic fractures often require fixation through the iliosacral joint, typically guided by fluoroscopy using an untracked C-arm device. However, this involves ionizing radiation exposure and potentially inaccurate screw placement. We introduce the Navigated Orthopaedic Fixations using Ultrasound System (NOFUSS), a radiation-free ultrasound (US)-based end-to-end system for providing real-time navigation for iliosacral screw (ISS) insertions.
Methods:
We performed surgeries on 8 human cadaver specimens, inserting four ISSs per specimen to directly compare NOFUSS against conventional fluoroscopy. Six specimens yielded usable (marginal or adequate quality) US images.
Results:
The median screw entry error, midpoint error, and angulations errors for NOFUSS were 8.4 mm, 7.0 mm, and 1.4, compared to 7.5 mm ( = 0.52), 5.7 mm ( = 0.30), and 4.4 ( = 0.001) for fluoroscopy respectively. NOFUSS resulted in 6 (50%) breaches, compared to 2 (16.7%) in fluoroscopy ( = 0.19). The median insertion time was 7 m 37 s and 12 m 36 s per screw for NOFUSS and fluoroscopy respectively ( = 0.002). The median radiation exposure during the fluoroscopic procedure was 2 m 44 s, (range: 1 m 44 s-3 m 18 s), with no radiation required for NOFUSS. When considering the three cadavers that yielded only adequate-quality US images (12 screws), the measured entry errors were 3.6 mm and 8.1 mm respectively for NOFUSS and fluoroscopy ( = 0.06).
Conclusion:
NOFUSS achieved insertion accuracies on par with the conventional fluoroscopic method, and reduced insertion times and radiation exposure significantly.
Significance:
This study demonstrated the feasibility of an automated, radiation-free, US-based surgical navigation system for ISS insertions.