A Detailed Protocol for Physiological Parameters Acquisition and Analysis in Neurosurgical Critical Patients
Published on: October 17, 2017
"Alexa, Cycle The Blood Pressure": A Voice Control Interface Method for Anesthesia Monitoring
Grace Lee1, Christopher W Connor1,2,3,4
1From the Department of Anesthesiology, Perioperative and Pain Medicine, Brigham and Women's Hospital-Harvard Medical School, Boston, Massachusetts.
Background:
Anesthesia monitors and devices are usually controlled with some combination of dials, keypads, a keyboard, or a touch screen. Thus, anesthesiologists can operate their monitors only when they are physically close to them, and not otherwise task-loaded with sterile procedures such as line or block placement. Voice recognition technology has become commonplace and may offer advantages in anesthesia practice such as reducing surface contamination rates and allowing anesthesiologists to effect changes in monitoring and therapy when they would otherwise presently be unable to do so. We hypothesized that this technology is practicable and that anesthesiologists would consider it useful.
Methods:
A novel voice-driven prototype controller was designed for the GE Solar 8000M anesthesia patient monitor. The apparatus was implemented using a Raspberry Pi 4 single-board computer, an external conference audio device, a Google Cloud Speech-to-Text platform, and a modified Solar controller to effect commands. Fifty anesthesia providers tested the prototype. Evaluations and surveys were completed in a nonclinical environment to avoid any ethical or safety concerns regarding the use of the device in direct patient care. All anesthesiologists sampled were fluent English speakers; many with inflections from their first language or national origin, reflecting diversity in the population of practicing anesthesiologists.
Results:
The prototype was uniformly well-received by anesthesiologists. Ease-of-use, usefulness, and effectiveness were assessed on a Likert scale with means of 9.96, 7.22, and 8.48 of 10, respectively. No population cofactors were associated with these results. Advancing level of training (eg, nonattending versus attending) was not correlated with any preference. Accent of country or region was not correlated with any preference. Vocal pitch register did not correlate with any preference. Statistical analyses were performed with analysis of variance and the unpaired t -test.
Conclusions:
The use of voice recognition to control operating room monitors was well-received anesthesia providers. Additional commands are easily implemented on the prototype controller. No adverse relationship was found between acceptability and level of anesthesia experience, pitch of voice, or presence of accent. Voice recognition is a promising method of controlling anesthesia monitors and devices that could potentially increase usability and situational awareness in circumstances where the anesthesiologist is otherwise out-of-position or task-loaded.
More Related Videos
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Assessment of blood pressure in brachial artery(two-step method)
Pre-Procedural Guidelines for Assessing Blood Pressure