Quality assurance and other challenges in paediatric radiotherapy: Accurate delivery of craniospinal radiotherapy

Alison L Salkeld1,2, Jonathan Sykes1,2, John Fernandez1

  • 1Department of Radiation Oncology, Sydney West Radiation Oncology Network, Western Sydney Local Health District, Sydney, New South Wales, Australia.

Insights

Setup variations in cranio-spinal radiotherapy (CSI) for paediatric and adolescent/young adult (AYA) patients had minimal dosimetric impact. Appropriate planning target volume (PTV) margins can compensate for occasional shifts, but systematic errors require regular review.

Area of Science:

  • Radiation Oncology
  • Paediatric Oncology
  • Medical Physics

Background:

  • Cranio-spinal radiotherapy (CSI) is crucial for central nervous system malignancies in paediatric, adolescent/young adult (AYA), and adult patients.
  • Delivering CSI in paediatric and AYA populations presents unique challenges across different age groups.
  • Accurate patient setup is vital for effective CSI treatment delivery.

Purpose of the Study:

  • To assess setup variations during CSI in paediatric and AYA patients.
  • To evaluate the dosimetric impact of these setup variations on target volumes and organs at risk (OARs).

Main Methods:

  • Retrospective analysis of 10 paediatric and AYA patients (aged 4-25) treated with volumetric modulated arc therapy (VMAT) CSI.
  • Quantification of setup errors by comparing daily treatment images with planned positions.
  • Evaluation of dosimetric impact on clinical target volumes (CTV) and OARs.

Main Results:

  • Setup errors ranged from 0.5 to 6.2 mm, particularly in the cranio-caudal direction.
  • Despite setup variations, minimal impact (<1% relative change) was observed on CTV coverage and OAR doses.
  • Occasional substantial setup variations had limited dosimetric consequences.

Conclusions:

  • Precise patient setup is essential for adequate CTV coverage in paediatric and AYA CSI.
  • Planning target volume (PTV) margins can effectively mitigate the impact of occasional setup variations.
  • Regular off-line review of setup trends is recommended to detect and address systematic errors, ensuring treatment quality.