Related Experiment Video
Updated: May 5, 2026

A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
Predictors of Shock-Reduction Programming and Its Impact on Implantable Cardioverter-Defibrillator Therapies and
Sebastian E Beyer1,2, Camden Harrell3, Steven Mullane3
1Division of Cardiology, Department of Medicine, Weill Cornell Medicine, New York- Presbyterian Hospital New York NY.
Background:
Shock-reduction implantable cardioverter-defibrillator programming (SRP) was associated with fewer therapies and improved survival in randomized controlled trials, but real-world studies investigating SRP and associated outcomes are limited.
Methods And Results:
The BIOTRONIK CERTITUDE registry was linked with the Medicare database. We included all patients with an implantable cardioverter-defibrillator implanted between August 22, 2012 and September 30, 2021 in the United States. SRP was defined as programming to either a therapy rate cutoff ≥188 beats per minute or number of intervals to detection ≥30/40 for treatment. Among 6781 patients (mean 74±9 years; 27% women), 3393 (50%) had SRP. Older age, secondary prevention indication, and device implantation in the southern or western United States were associated with lower use of SRP. The cumulative incidence rate of implantable cardioverter-defibrillator shocks was lower in the SRP group (5.1% shocks/patient year) compared with the non-SRP group (7.2% shocks/patient year) (adjusted hazard ratio [HR], 0.83 [95% CI, 0.73-0.96]; P=0.005). Over a median follow-up of 2.9 years, 739 deaths occurred in the SRP group and 822 deaths occurred in the non-SRP group (adjusted HR, 0.97 [95% CI, 0.88-1.07]; P=0.569). SRP was associated with a lower all-cause mortality among patients without ischemic heart disease compared with patients with ischemic heart disease (adjusted HR, 0.64 [95% CI, 0.48-0.87] versus adjusted HR, 1.02 [95% CI, 0.92-1.14]; Pinteraction=0.004).
Conclusions:
Adoption of SRP is low in real-world clinical practice. Age, clinical variables, and geographic factors are associated with use of SRP. In this study, SRP-associated decrease in mortality was limited to patients without ischemic heart disease.
More Related Videos
12:45Benefits of Cardiac Resynchronization Therapy in an Asynchronous Heart Failure Model Induced by Left Bundle Branch Ablation and Rapid Pacing
Published on: December 11, 2017
06:10Utilizing Percutaneous Ventricular Assist Devices in Acute Myocardial Infarction Complicated by Cardiogenic Shock
Published on: June 12, 2021
Related Concept Videos
Coronary Artery Disease IV: Preventive Measures
Acute Coronary Syndrome III: Diagnostic Studies
Cardiopulmonary Resuscitation III: AED Use
Cardiomyopathy II: Dilated Cardiomyopathy
Cardiomyopathy V: Interprofessional Care