Objective:
The aim of this study was to compare the Morse taper (MT) + titanium base (Ti-Base) abutment with the external hexagon (EH) + Ti-Base abutment by using the strain gauge method in the mesial, distal, and apical-buccal areas around these types of implants.
Materials And Methods:
This study investigated two groups, MT and EH, each comprising five polyurethane samples with a dental implant (3.75 × 11.5 mm) in the area of artificial tooth 15. The strain gauges were glued to the mesial, distal, and apical-buccal polyurethane areas of all samples in relation to the implant. Ti-Base nonangled abutments were installed on the implants in each group. Ten identical zirconia crowns were constructed by scanning and milling and were subsequently cemented onto the Ti-base abutments with calcium hydroxide cement. Then, an axial load of 100 N was applied to the occlusal region of the zirconia crowns, and strain gauge measurements were taken.
Statistical Analysis:
Strain gauge data were assessed by a two-way analysis of variance (ANOVA) with "implant connection" and "strain gauge position" factors, followed by the Bonferroni test (p < 0.05).
Results:
The MT group showed significantly lower microstrain values in the mesial and apical strain gauges compared to the EH group.
Conclusion:
The MT group exhibited less microstrain in the mesial and apical areas of the polyurethane samples near the implant. Consequently, the MT connection was considered more biomechanically advantageous.
More Related Videos
07:35Creating Rigidly Stabilized Fractures for Assessing Intramembranous Ossification, Distraction Osteogenesis, or Healing of Critical Sized Defects
Published on: April 11, 2012
11:38Environmentally-controlled Microtensile Testing of Mechanically-adaptive Polymer Nanocomposites for ex vivo Characterization
Published on: August 20, 2013
Related Concept Videos
Residual Stresses in Circular Shafts
Three-Dimensional Analysis of Strain
Mohr's Circle for Plane Strain
Mohr's circle visually represents the strain states under various conditions, which is essential for...
Plastic Deformation in Circular Shafts
Stress Concentrations in Circular Shafts
Measurements of Strain