Related Experiment Video
Updated: Jun 21, 2025

Surfactant Depletion Combined with Injurious Ventilation Results in a Reproducible Model of the Acute Respiratory Distress Syndrome ARDS
Published on: April 7, 2021
The Impact of PEEP on Ventilation Distribution in ARDS
Bruno Louis1, Martin Cour2, Laurent Argaud2
1Institut Mondor de Recherches Biomédicales INSERM-UPEC U955 CNRS EMR7000, Créteil, France. bruno.louis@inserm.fr.
Background:
The first aim of this study was to evaluate the capacity of electrical impedance tomography (EIT) to identify the effect of PEEP on regional ventilation distribution and the regional risk of collapse, overdistention, hypoventilation, and pendelluft in mechanically ventilated patients. The second aim was to evaluate the feasibility of EIT for estimating airway opening pressure (AOP).
Methods:
The EIT signal was recorded both during baseline cyclic ventilation and slow insufflation for one breath for 9 subjects with moderate-to-severe ARDS. From these data, the AOP and volumes insufflated to lung regions with or without the risk of either collapse, overdistention, hypoventilation, or pendelluft were assessed at 3 PEEP levels (5, 10, and 15 cm H2O). PEEP levels were compared by Friedman analysis of variance and the AOP measured by EIT evaluated using an F-test and the Bland and Altman method.
Results:
The volume for which there was no specific risk significantly decreased at the highest PEEP from 55 ± 31% tidal volume (VT) at PEEP 5 or 82 ± 18% VT at PEEP 10 to 10 ± 30% VT at PEEP 15 (P = .038 between PEEP 5 vs PEEP 15; P = .01 between PEEP 10 vs PEEP 15). The volume associated with overdistention significantly increased with increasing PEEP, whereas that associated with atelectrauma significantly decreased. Pendelluft significantly decreased with increasing PEEP: VT of 8.9 ± 18.6%, 3.6 ± 7.0%, and 3.2 ± 7.1% for PEEP 5, PEEP 10, and PEEP 15, respectively. The center of ventilation tended to increase in the dependent direction with higher PEEP. The AOPs assessed by EIT and from the pressure-volume curve were in good agreement (bias 0.48 cm H2O).
Conclusions:
Our results suggest that EIT could aid clinicians in making personalized and reasoned choices in setting the PEEP for subjects with ARDS.
Related Concept Videos
Factors Affecting Pulmonary Ventilation
Alveolar Surface Tension
The alveolar fluid lines the luminal surface of the alveoli and exerts a force called surface tension. This force is caused by the polar water molecules in the liquid being more strongly attracted to each...
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Mechanical Ventilation I: Indication and Settings
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...