An investigation of tomosynthesis on the diagnostic efficacy of spot compression mammography
Ying Jiang1, Lilin Yang1, Rong Qian1
1Department of Radiology, Sichuan Provincial People's Hospital, School of Medicine, University of Electronic Science and Technology of China, Chengdu, China.
Abstract:
To explore the diagnostic efficacy of tomosynthesis spot compression (TSC) compared with conventional spot compression (CSC) for ambiguous findings on full-field digital mammography (FFDM). In this retrospective study, 122 patients (including 108 patients with dense breasts) with ambiguous FFDM findings were imaged with both CSC and TSC. Two radiologists independently reviewed the images and evaluated lesions using the Breast Imaging Reporting and Data System. Pathology or at least a 1-year follow-up imaging was used as the reference standard. Diagnostic efficacies of CSC and TSC were compared, including area under the curve (AUC), accuracy, sensitivity, specificity, positive predictive value (PPV), and negative predictive value (NPV). The mean glandular dose was recorded and compared for TSC and CSC. Of the 122 patients, 63 had benign lesions and 59 had malignant lesions. For Reader 1, the following diagnostic efficacies of TSC were significantly higher than those of CSC: AUC (0.988 vs. 0.906, P = 0.001), accuracy (93.4% vs. 77.8%, P = 0.001), specificity (87.3% vs. 63.5%, P = 0.002), PPV (88.1% vs. 70.5%, P = 0.010), and NPV (100% vs. 90.9%, P = 0.029). For Reader 2, TSC showed higher AUC (0.949 vs. 0.909, P = 0.011) and accuracy (83.6% vs. 71.3%, P = 0.022) than CSC. The mean glandular dose of TSC was higher than that of CSC (1.85 ± 0.53 vs. 1.47 ± 0.58 mGy, P < 0.001) but remained within the safety limit. TSC provides better diagnostic efficacy with a slightly higher but tolerable radiation dose than CSC. Therefore, TSC may be a candidate modality for patients with ambiguous findings on FFDM.
More Related Videos
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Radiological Investigation I: X-ray and CT