Computer-assisted revision total knee arthroplasty does not improve postoperative knee prosthesis alignment compared
Triine E Alling1, Marrigje F Conteh-Meijer1,2, Alexander L Boerboom1
1Department of Orthopaedics, University Medical Center Groningen University of Groningen Groningen The Netherlands.
Computer-assisted surgery (CAS) did not improve prosthesis alignment in revision total knee arthroplasty (rTKA). This study found no significant differences in coronal, sagittal, or rotational alignment between CAS-rTKA and conventional rTKA.
Area of Science:
- Orthopedic surgery
- Surgical technology
- Biomedical engineering
Background:
- Computer-assisted surgery (CAS) is established for primary total knee arthroplasty (TKA) prosthesis alignment.
- Literature on CAS use in revision TKA (rTKA) is limited.
- The impact of CAS on rotational alignment in rTKA remains undescribed.
Purpose of the Study:
- To evaluate the effect of CAS in rTKA.
- To compare coronal, sagittal, and rotational prosthesis alignment between CAS-assisted rTKA and conventional rTKA.
- To assess differences in outlier proportions between the two techniques.
Main Methods:
- Prospective cohort study comparing CAS-rTKA (54 patients/62 knees) with a historical control group (CON-rTKA) (13 patients/23 knees).
- CAS-rTKA utilized imageless CAS between 2012-2017.
- Conventional rTKA (CON-rTKA) used standard techniques between 2002-2012.
- Postoperative alignment assessed via EOS-2D/3D (coronal/sagittal) and CT scans (rotation).
Main Results:
- No significant differences in coronal or sagittal alignment (e.g., mechanical leg angle, distal femoral/proximal tibial angles) between CAS-rTKA and CON-rTKA groups.
- No significant differences in femoral or tibial rotation between the groups.
- The proportion of outliers in coronal, sagittal, and rotational alignment did not differ significantly between the CAS-rTKA and CON-rTKA groups.
Conclusions:
- This study found no evidence that CAS improves coronal, sagittal, or rotational alignment in rTKA.
- No significant difference in outlier rates was observed between CAS-assisted and conventional rTKA.
- Current evidence does not support the routine use of CAS for enhancing prosthesis alignment in revision TKA.