Rapid prototyping of a retinal multivascular network phantom for optical retinal vascular imaging equipment
Xiaowei Zhao1, Wenli Liu1, Zhixiong Hu1
1Center for Medical Metrology, National Institute of Metrology, 18 North 3rd Ring East Rd., Chaoyang, Beijing 100029, China.
Abstract:
Retinal vascular health holds paramount importance for healthy vision. Many technologies have been developed to examine retinal vasculature non-destructively, including fundus cameras, optical coherence tomography angiography (OCTA), fluorescein angiography (FA), and so on. However, there is a lack of a proper phantom simulating the critical features of the real human retina to calibrate and evaluate the performance of these technologies. In this work, we present a rapid, high-resolution, and economical technology based on 3D printed mold-based soft lithography and spin coating for the fabrication of a multivascular network and multilayer structural retinal phantom with the appropriate optical properties. The feasibility of the retinal phantom as a test device was demonstrated with an OCTA system and a confocal retinal ophthalmoscope. Experiment results prove that the retinal phantom could provide an objective evaluation of the OCTA and confocal retinal ophthalmoscope. Furthermore, the microfluidic phantoms enabled by this fabrication technology may support the development and evaluation of other techniques.