Validation of Acuros for total body irradiation at extended distance
Geoffrey Nelson1, Vikren Sarkar1, Martin Szegedi1
1Department of Radiation Oncology, University of Utah, Salt Lake City, Utah, USA.
Purpose:
Standardized and accurately reported doses are essential in conventional total body irradiation (TBI), especially lung doses. This study evaluates the accuracy of the Acuros algorithm in predicting doses for extended-distance TBI.
Methods:
Measurements and calculations were done with both 6 and 18 MV. Tissue Maximum Ratio (TMR), output and off axis ratios (OAR) were measured at 200 and 500 cm source to detector distance and compared to Acuros calculated values. Two end-to-end tests were carried out, one with an in-house phantom (solid water and Styrofoam) with inserted ion chambers and the other was with the Imaging and Radiation Oncology Core (IROC) TBI anthropomorphic phantom equipped with TLDs. The end-to-end test was done for 6 and 18 MV both with and without lung blocks. The source to midplane distance for both phantoms were at 518 and 508 cm respectively. Lung blocks were placed at the phantom surface and a beam spoiler was positioned 30 cm from the surface of the phantoms as per our clinical set up.
Results:
The agreement between measured and calculated TMR, output and off axis ratios for both 6 and 18 MV were within 2%. Ion chamber measurements in both the Styrofoam and solid water for both energies carried out with and without lung blocks were within 2% of calculated values. TLD measured doses for both 6 and 18 MV in the IROC phantom were within 5% of calculated doses which is within the uncertainty of the TLD measurement.
Conclusions:
The results indicate that the clinical beam model for Acuros 16.1 commissioned at standard clinical distances is capable of calculating doses accurately at extended distances up to 500 cm.
More Related Videos
08:54Treatment of Liver Metastases Using an Internal Target Volume Method for Stereotactic Body Radiotherapy
Published on: May 8, 2018
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
Related Concept Videos
Radiation: Applications
The average...
Absorption of Radiation