Related Experiment Video
Updated: Jun 20, 2025

Insertion, Maintenance, and Removal of the Percutaneous Dual Lumen Cannula Right Ventricular Assist Device
Published on: July 20, 2022
The double barrel Impella exchange: A reliable method for uninterrupted mechanical circulatory support
Jessica S Clothier1, Serge Kobsa1, Jonathan Praeger1
1Division of Cardiac Surgery, University of Southern California Keck School of Medicine, Los Angeles, California.
Abstract:
Classic Impella exchange interrupts flow when the old device is pulled into the aorta before advancing the new device across the aortic valve, threatening circulatory collapse and loss of left ventricular access. In "double barrel," uninterrupted Impella exchange, the new device is placed into the ventricle alongside the old, where flow is first transitioned completely. Of 31 consecutive patients undergoing this procedure, none experienced intraoperative cardiac arrest, and 96.8% (30/31) had no new aortic insufficiency. One vascular complication ensued following known preoperative iliac injury. One patient suffered nonembolic stroke; another had subarachnoid hemorrhage. Fifty-five percent (17/31) of patients survived, with 22.6% (7/31) recovering, 25.8% (8/31) undergoing transplant, and 6.5% (2/31) transitioning to durable left ventricular assist device. Impella-only survival (83.3%, 10/12) was significantly higher than Impella-extracorporeal membrane oxygenation survival (36.8%, 7/19; (operating room) OR 14.46, 95% ((confidence interval) CI 1.74-119.93, p = 0.01). We conclude the "double barrel" technique is reliable in device-dependent cardiogenic shock patients, offering significant advantages and minimal risk.