Related Experiment Video
Updated: Jun 20, 2025

08:51
Monitoring Protein Adsorption with Solid-state Nanopores
Published on: December 2, 2011
13.5K
Immunological assay using a solid-state pore with a low limit of detection.
Hiroyasu Takei1, Tomoko Nakada2, Lat Wai Leong2
1Aipore Inc., 26-1 Sakuragaokacho, Shibuya, Tokyo, 150-8512, Japan.
Scientific Reports
|July 19, 2024
Summary
This study presents a novel method combining immunoreactions and solid-state pores for highly sensitive detection of tumor markers. This breakthrough enables rapid, quantitative point-of-care testing with a significantly lower detection limit.
Area of Science:
- Biotechnology
- Nanotechnology
- Immunology
Background:
- Immune reactions are crucial for disease detection, relying on antigen-antibody recognition.
- Solid-state pores offer single-molecule detection by varying pore diameter.
- Combining these technologies promises sensitive and specific disease testing.
Purpose of the Study:
- To develop a quantitative analysis method using solid-state pores for disease detection.
- To achieve a low limit of detection for tumor markers via hybrid technology.
- To create rapid, sensitive, and specific immunoassay tests.
Main Methods:
- Utilized a hybrid approach combining immunoreactions with solid-state pore technology.
- Employed prostate-specific antigen (PSA) and anti-PSA antibody-modified beads.
- Demonstrated quantitative analysis of bead mixtures and PSA detection.
Main Results:
- Achieved quantitative analysis of bead mixtures with an error rate up to 4.7%.
- Reached a detection limit of 24.9 fM for prostate-specific antigen (PSA) in 30 minutes.
- Demonstrated a lower limit of detection and higher throughput compared to existing immunoassays.
Conclusions:
- The hybrid solid-state pore and immunoreaction method enables sensitive tumor marker detection.
- This technology facilitates the development of rapid, easy-to-use point-of-care tests.
- The method offers improved sensitivity and throughput for immunoassay applications.