Automatic vessel attenuation measurement for quality control of contrast-enhanced CT: Validation on the portal vein
Kevin McCoy1,2, Sujay Marisetty3, Dominique Tan4
1Department of Statistics, Rice University, Houston, Texas, USA.
Background:
Adequate image enhancement of organs and blood vessels of interest is an important aspect of image quality in contrast-enhanced computed tomography (CT). There is a need for an objective method for evaluation of vessel contrast that can be automatically and systematically applied to large sets of CT exams.
Purpose:
The purpose of this work was to develop a method to automatically segment and measure attenuation Hounsfield Unit (HU) in the portal vein (PV) in contrast-enhanced abdomen CT examinations.
Methods:
Input CT images were processed by a vessel enhancing filter to determine candidate PV segmentations. Multiple machine learning (ML) classifiers were evaluated for classifying a segmentation as corresponding to the PV based on segmentation shape, location, and intensity features. A public data set of 82 contrast-enhanced abdomen CT examinations was used to train the method. An optimal ML classifier was selected by training and tuning on 66 out of the 82 exams (80% training split) in the public data set. The method was evaluated in terms of segmentation classification accuracy and PV attenuation measurement accuracy, compared to manually determined ground truth, on a test set of the remaining 16 exams (20% test split) held out from public data set. The method was further evaluated on a separate, independently collected test set of 21 examinations.
Results:
The best classifier was found to be a random forest, with a precision of 0.892 in the held-out test set to correctly identify the PV from among the input candidate segmentations. The mean absolute error of the measured PV attenuation relative to ground truth manual measurement was 13.4 HU. On the independent test set, the overall precision decreased to 0.684. However, the PV attenuation measurement remained relatively accurate with a mean absolute error of 15.2 HU.
Conclusions:
The method was shown to accurately measure PV attenuation over a large range of attenuation values, and was validated in an independently collected dataset. The method did not require time-consuming manual contouring to supervise training. The method may be applied to systematic quality control of contrast-enhanced CT examinations.
More Related Videos
06:46Quantitative Micro-CT Analysis of Aortopathy in a Mouse Model of β-aminopropionitrile-induced Aortic Aneurysm and Dissection
Published on: July 16, 2018
08:41Novel In Vivo Micro-Computed Tomography Imaging Techniques for Assessing the Progression of Non-Alcoholic Fatty Liver Disease
Published on: March 24, 2023
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Venous Thrombosis II: Clinical Manifestations and Diagnostic Studies
Imaging Studies VII: Vascular Imaging