Related Experiment Video
Updated: May 5, 2026

A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
Cardiac resynchronization therapy-defibrillator implantation with shock lead placement in the left bundle branch
Kenta Yoshida1, Mitsuru Yoshino1, Tokuma Kawabata1
1Department of Cardiovascular Medicine, Kurashiki Central Hospital, 1-1-1 Miwa, Kurashiki 710-8602, Japan.
Background:
Cardiac resynchronization therapy (CRT) with biventricular pacing is a well-established therapy. Left bundle branch area pacing (LBBAP) is a safe technique providing physiological pacing, and LBBAP-optimized CRT (LOT-CRT) has been shown to provide better electrical resynchronization than traditional CRT. However, there are few reports on shock lead placement in the left bundle branch area (LBBA) during CRT-defibrillator (CRT-D) implantation.
Case Summary:
A 76-year-old woman with heart failure from dilated cardiomyopathy presented with left bundle branch block pattern (QRS duration, 160 ms). Left ventricular ejection fraction was 21%. Cardiac resynchronization therapy-defibrillator implantation was performed due to worsening symptoms. By reshaping the Agilis HisPro catheter and adding a septal curve, the shock lead was placed deep into the ventricular septum, narrowing QRS duration to 114 ms. Left ventricular activation time was 84 ms. A defibrillation threshold test confirmed successful treatment without adverse events. At 6-month follow-up, left ventricular ejection fraction improved from 21 to 63%, with the patient's condition improving from New York Heart Association class III to class I.
Discussion:
It was reported that QRS narrowing in CRT was related to long-term mortality, and LOT-CRT further decreased QRS duration as compared with LBBP only or biventricular pacing and increased the response rate. Combining LBBAP with coronary sinus pacing can potentially achieve superior electrical resynchronization. Lack of a suitable tool for direct shock lead placement in LBBA necessitated additional LBBAP lead in conventional LOT-CRT. Our successful LOT-CRT-D procedure with minimal number of leads through Agilis HisPro catheter reshaping enabled direct LBBA shock lead placement.
More Related Videos
12:45Benefits of Cardiac Resynchronization Therapy in an Asynchronous Heart Failure Model Induced by Left Bundle Branch Ablation and Rapid Pacing
Published on: December 11, 2017
10:17Real-Time Cardiac Mapping with a Noninvasive Imageless Electrocardiographic Imaging System
Published on: April 11, 2025
Related Concept Videos
Dysrhythmias VI: Management of Dysrhythmias
Cardiomyopathy II: Dilated Cardiomyopathy
Cardiomyopathy V: Interprofessional Care