Double testing with lateral flow antigen test devices for COVID-19: does a second test in quick succession add value?
Matthias E Futschik1, Sarah A Tunkel2, Elena Turek3
1UK Health Security Agency, London, UK; School of Biomedical Sciences, Faculty of Health, University of Plymouth, Plymouth, UK.
Performing two rapid lateral flow device (LFD) tests for severe acute respiratory syndrome-related coronavirus 2 (SARS-CoV-2) antigen detection showed high reproducibility. However, this method did not yield statistically significant improvements in diagnostic sensitivity or specificity compared to a single test.
Area of Science:
- Infectious Diseases
- Diagnostic Technologies
- Public Health
Background:
- Severe acute respiratory syndrome-related coronavirus 2 (SARS-CoV-2) antigen detection relies on lateral flow devices (LFDs).
- Optimizing LFD performance is crucial for accurate and timely diagnosis.
- Assessing novel testing strategies can enhance diagnostic capabilities.
Purpose of the Study:
- To evaluate if performing two sequential LFD tests improves diagnostic sensitivity and specificity for SARS-CoV-2 antigen detection.
- To determine the impact of a confirmatory LFD test on overall diagnostic accuracy.
- To assess the reproducibility of LFD tests performed by untrained users.
Main Methods:
- A study involving individuals aged 16 years and older at UK community testing sites (February-May 2021).
- Participants performed two successive LFD tests, followed by a polymerase chain reaction (PCR) test as the reference standard.
- Sensitivity was assessed by considering either LFD result as positive; specificity was evaluated using the second LFD as a confirmatory test.
Main Results:
- High concordance (>97%) was observed between sequential LFD results and between LFDs and PCR tests.
- Combined LFD usage slightly increased sensitivity (68.6%) and specificity (99.8%) compared to individual tests, but these increases were not statistically significant.
- LFD tests demonstrated high reproducibility, even when performed by untrained users without supervision.
Conclusions:
- Lateral flow device tests are highly reproducible for SARS-CoV-2 antigen detection, even in unsupervised settings.
- Performing two LFD tests in rapid succession offers marginal improvements in sensitivity and specificity.
- Statistically significant enhancements in diagnostic accuracy were not achieved by using two sequential LFD tests in this study.