Background:
When cardiac implantable electronic device infection occurs, standard therapy is usually total system extraction. Transvenous lead extraction is preferable to open heart surgical extraction, unless contraindicated because of the presence of very large vegetations on the intravenous leads according to the European Society of Cardiology guidelines. Extraction of transvenous leads with vegetations risks distal embolism resulting in obstruction and/or infection in the pulmonary arteries. Catheter aspiration of vegetations or thrombi has been performed prior to transvenous lead extraction using a partial veno-venous extracorporeal bypass circuit. We report the use of a single-access aspiration system using the Inari FlowTriever 24 French system to debulk a defibrillator lead before percutaneous extraction.
Case Summary:
A 79-year-old male presented with fever 18 years after his first implantable cardioverter defibrillator implant and 9 years after his most recent pulse generator change. Two large vegetations were identified on his transvenous defibrillator lead on the atrial aspect, near the tricuspid annulus, which were aspirated using the Inari Medical 24Fr FlowTriever aspiration catheter. We describe anatomical considerations during the approach and a technique to localize the vegetations based on a combination of fluoroscopy and transoesophageal echocardiogram guidance.
Discussion:
This case demonstrates the safe and effective use of the Inari Medical 24Fr FlowTriever aspiration catheter in debulking a defibrillator lead before transvenous lead extraction. This method uses a single venous puncture and is not dependent on extracorporeal bypass. Apart from reducing complexity, this technique may be advantageous in patients where anticoagulation needs to be minimised.
More Related Videos
10:46Non-fluoroscopic Catheter Tracking for Fluoroscopy Reduction in Interventional Electrophysiology
Published on: May 26, 2015
12:45Benefits of Cardiac Resynchronization Therapy in an Asynchronous Heart Failure Model Induced by Left Bundle Branch Ablation and Rapid Pacing
Published on: December 11, 2017
Related Concept Videos
Cardiac Catheterization I: Pre-Procedure Overview
Cardiac Catheterization II: Right Heart Catheterization
Cardiac Catheterization III: Left Heart Catheterization
Cardiac Catheterization IV: Nursing Management