Stroke Evaluation in the Interventional Suite Using Dual-Layer Detector Cone-Beam CT: a First-in-human Prospective
Fredrik Ståhl1,2, Håkan Almqvist3,4, Åsa Aspelin3
1Department of Neuroradiology, Karolinska University Hospital, Eugeniavägen 3, Stockholm, Sweden. Fredrik.stahl@regionstockholm.se.
Purpose:
Cone-beam CT in the interventional suite could be an alternative to CT to shorten door-to-thrombectomy time. However, image quality in cone-beam CT is limited by artifacts and poor differentiation between gray and white matter. This study compared non-contrast brain dual-layer cone-beam CT in the interventional suite to reference standard CT in stroke patients.
Methods:
A prospective single-center study enrolled consecutive participants with ischemic or hemorrhagic stroke. The hemorrhage detection accuracy, per-region ASPECTS accuracy and subjective image quality (Likert scales for gray-white matter differentiation, structure perception and artifacts) were assessed by three neuroradiologists blinded to clinical data on dual-layer cone-beam CT 75 keV monoenergetic images compared to CT. Objective image quality was assessed by region-of-interest metrics. Non-inferiority for hemorrhage detection and ASPECTS accuracy was determined by the exact binomial test with a one-sided lower performance boundary prospectively set to 80% (98.75% CI).
Results:
27 participants were included (74 years ± 9; 19 female) in the hyperacute or acute stroke phase. One reader missed a small bleeding, but all hemorrhages were detected in the majority analysis (100% accuracy, CI lower boundary 86%, p = 0.002). ASPECTS majority analysis showed 90% accuracy (CI lower boundary 85%, p < 0.001). Sensitivity was 66% (individual readers 67%, 69%, and 76%), specificity was 97% (97%, 96%, 89%). Subjective and objective image quality were inferior to CT.
Conclusion:
In a small single-center cohort, dual-layer cone-beam CT showed non-inferior hemorrhage detection and ASPECTS accuracy to CT. Despite inferior image quality, the technique may be useful for stroke evaluation in the interventional suite.
Trial Registration Number:
NCT04571099 (clinicaltrials.gov). Prospectively registered 2020-09-04.
More Related Videos
09:49Dual-phase Cone-beam Computed Tomography to See, Reach, and Treat Hepatocellular Carcinoma during Drug-eluting Beads Transarterial Chemo-embolization
Published on: December 2, 2013
06:59Improved Registration of 3D CT Angiography with X-ray Fluoroscopy for Image Fusion During Transcatheter Aortic Valve Implantation
Published on: June 3, 2018
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Radiological Investigation I: X-ray and CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies III: Computed Tomography